Siemens SITRANS WF100 Operating Instructions Manual page 31

Solid flowmeters
Hide thumbs Also See for SITRANS WF100:
Table of Contents

Advertisement

4. After the span calibration has been completed, remove the test weight and pulley and store
them.
Material test
Calibration of the flowmeter by test weight is not a dynamic simulation and yields only an
approximate span calibration. Effective material flow, friction, and velocity can only be
reproduced by running actual material tests.
At least two 5 minute samples should be taken to ensure repeatability. Refer to material test and
subsequent factoring of the flowmeter sections of the integrator manual.
Re-rating
Consult your local Siemens Milltronics representitive for re-rating.
SITRANS WF100
Operating Instructions, 02/2012, 7ML19985NB01-01
SITRANS WF100 housing
Hook
Calibration access hole (remove plug to access hook during calibration)
Retaining pin
Calibration pulley
Sensing plate
Calibration bracket
Cable
Test weight
Commissioning