2.9 Menu

Pressing the MENU button will display the first page of
the MENU screen.
Set Pass Levels
Set Auto/Manual
Recall Memory
Scrolling down with the i button will display the second
page of the MENU screen.
Delete All Memory
Reset to Defaults
Use the f button to exit the MENU screen.
2.10 Null of Test Probe Lead Resistance (Class I
Earth Continuity only)
The null function is used to compensate for the test
probe lead resistance when used for Class I earth
continuity measurements. This also allows for the use
of alternative test probes.
To null the test probe lead resistance press SHIFT and
NULL (LOAD) to display the NULL screen. Use hi to
select Null or Abort.
If Abort is selected, press OK and the HPAT600 will
return to the SELECT screen.
If Null is selected, connect the test probe from the
4mm socket (2) to the Earth pin of the front panel
mains socket (1) (use the brass earth contact pin
supplied with the HPAT600). While connected, press
OK.
Refer to 2.11
Refer to 2.13
Refer to 2.16
Refer to 2.19
Refer to 2.20