ABB AX400 Series User Manual page 70

Single and dual input analyzers for ph/redox (orp)
Hide thumbs Also See for AX400 Series:
Table of Contents

Advertisement

Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
Temp
ºC
ºF
4.01 pH
0
32
4.003
5
41
3.998
10
50
3.996
15
59
3.996
20
68
3.999
25
77
4.004
30
86
4.011
35
95
4.020
40
104
4.030
45
113
4.042
50
122
4.055
55
131
4.070
60
140
4.085
65
149
70
158
4.120
75
167
80
176
4.160
85
185
90
194
4.190
95
203
4.210
Table A.4 NIST Buffer Solutions
68
NIST Buffers
6.86 pH
9.18 pH
6.982
9.460
6.949
9.392
6.921
9.331
6.898
9.276
6.878
9.227
6.863
9.183
6.851
9.143
6.842
9.107
6.836
9.074
6.832
9.044
6.831
9.017
Temp
ºC
ºF
4.01 pH
0
32
4.000
5
41
3.998
10
50
3.997
15
59
3.998
20
68
4.001
25
77
4.005
30
86
4.011
35
95
4.018
40
104
4.027
45
113
4.038
50
122
4.050
55
131
4.064
60
140
4.080
65
149
4.097
70
158
4.116
75
167
4.137
80
176
4.159
85
185
4.183
90
194
4.208
95
203
4.235
Table A.5 US Technical Buffer Solutions
Appendix A – Buffer Solutions
US Technical Buffers
7 pH
10.01 pH
7.118
10.317
7.087
10.245
7.059
10.179
7.036
10.118
7.016
10.062
7.000
10.012
6.987
9.966
6.977
9.925
6.970
9.889
6.965
9.857
6.964
9.828
6.965
6.968
6.974
6.982
6.992
7.004
7.018
7.034
7.052
IM/AX4PH Rev. P

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax416Ax466Ax468Ax436Ax460

Table of Contents