Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
-----
4.O1pH
Set Buffer 1
-----
9.18pH
Set Buffer 2
Set Auto Buffers
20
Set Buffer 1
Set the pH value of the buffer 1 solution – see Appendix A for pH tables.
Set Buffer 2
Set the pH value of the buffer 2 solution.
Note. The solution selected for buffer 2 must be at least 2 ph greater than that
selected for buffer 1, e.g. if buffer 1 is set to 7 pH, buffer 2 must be set to at least 9 pH.