Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
150. 0
Ohms
150. 0
Calib
A:T.Span (150R)
1000
Ohms
1000
Calib
A:T.Zero (1K0)
1500
Ohms
1500
Calib
A:T.Span (1K5)
2000
Ohms
2000
Calib
A:T.Zero (2K0)
5000
Ohms
5000
Calib
A:T.Span (5K0)
Cal. Sensor A
No
-----
Yes
A: Abort Cal.
IM/AX4PH Rev. P
Temperature Span (150R)
Set the temperature simulator to 150 
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (1k0)
Set the temperature simulator to 1000 
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (1k5)
Set the temperature simulator to 1500 
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (2k0)
Set the temperature simulator to 2000 
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (5k0)
Set the temperature simulator to 5000 
The display returns automatically to
recorded.
Abort Calibration
Select Yes or No
selected:
Yes
completion of
before
A:T.Zero (100R)
and continues.
(100R)
completion of
CAL. SENSOR A
after
page.
Sensor A
selected – calibration continues from the point at which the
No
pressed.
Cal. Sensor A
frame – calibration advances to
A:mV Span (+1V)
frame – the display returns to the
A:mV Span (+1V)
7 Calibration
once a stable and valid value is
A:T.Zero
Calibrate
key was
59