Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
A: Probe Type
pH
Antim.
-----
Glass
A: Electrode
Manual
-----
Auto
A: Temp. Comp.
Manual
Auto
3K Bal
Pt1000
-----
Pt100
A: Temp. Sensor
25. 0
Deg.C
A: Preset Temp.
IM/AX4PH Rev. P
pH Electrode Type
Select the type of pH electrode used, Glass or Antimony.
Temperature Compensation
Select
to enable the analyzer to compensate automatically for fluctuations in
Auto
sample temperature.
Temperature Sensor (Automatic Temperature Compensation only)
Select the type of temperature sensor used: Pt100, Pt1000 or Balco 3K.
continued on next page.
A: Sample Comp.
Preset Temperature (Manual Temperature Compensation only)
Enter the temperature of the sample within the range –10.0 to 120.0 ºC.
continued on next page.
A: Set Min Slope
5 Programming
31