W A R N I N G !
15

MAINTENANCE

Before carrying out any maintenance work, the specific safety instructions referring to the instalment
and the process must be observed!
W A R N I N G !
The frequency of the maintenance work depends on the operational process and can therefore only
be determined in each individual case.
All parts which require maintenance work are easily accessible and are installed on the front side of
the gas conditioning unit CSS-M/W. Only for any work on the gas sample pump, the case has to be
opened:
 Replace the filter element of the fine filter FPF-2-0,3 GF in case the flow rate is to low or after
visible control.
N O T E !
 The hose of the condensate pump SR25.2 should be checked every six months and, if necessary,
replaced (see chapter 15.3.1);
 The diaphragm of the sample gas pump N3 KPE should be checked every six months and, if
necessary, replaced.
www.mc-techgroup.com
Aggressive condensate possible.
Wear protective glasses and proper protective clothing!
Dangerous voltage. Before any maintenance work is carried out on
the gas conditioning system, put the main switch of the instrument
into position "0" and tear the mains plug off!
In order to protect downstream analysers in case of a irruption of
condensate, always replace the humid filter element.
CSS-M/W | 1.02.00