Supply Connections; Hose Connections; Figure 3 Hose Connections Css-M/W - M&C CSS Series Instruction Manual

Wall mount gas conditioning unit mount
Hide thumbs Also See for CSS Series:
Table of Contents

Advertisement

12

SUPPLY CONNECTIONS

12.1

HOSE CONNECTIONS

Figure 3
Hose connections CSS-M/W
Do not mix up the hose connections; the connections are marked
accordingly in figure 3.
N O T E !
After having connected all lines, please check the seals.
All hose connections are DN4/6 mm sealing ring threaded hose couplings out of PVDF/PPH as
standard. They are suitable for gas inlet temperatures of maximum 80 °C (176 °F) (see chapter 8).
The mounting of the sample gas hoses and the condensate hoses is to be executed as follows:
 Remove the union nut from the sealing ring couplings by turning it anti-clockwise. The nut should
be removed from the thread with great care so as to ensure that the loose sealing ring in the nut is
not lost.
 Place the union nut over the connecting hose.
 Push the sealing ring over the connecting hose with the thicker bead towards the nut.
 Place the hose over the nipple on the thread.
The tightness of the connections can only be guaranteed if the connecting
hose has a straight edge (use a hose cutter).
N O T E !
 The union nut is to be screwed tight by hand.
www.mc-techgroup.com
Sample gas inlet DN4/6
Mains cable
3000mm
Condensate out DN3,2/6,4
350mm
CSS-M/W | 1.02.00
Sample gas outlet DN4/6
Signal cable
3000mm
11