The hose will no longer be able to slip off, and is now compression-proof.
The hoses are to be removed in the reverse order.
W A R N I N G !
12.2

ELECTRICAL CONNECTIONS

W A R N I N G !
N O T E !
The gas conditioning unit CSS-M/W is available with either 230 V / 50 Hz or 115 V / 60 Hz (for circuit
diagram see Appendix). A 4 A fuse is used on all models as fuse protection. The fuse is located on
the electronic board inside the steel sheet case.
The electrical connection is made via the mains cable of 3 x 1 mm
side of the case.
Brown
Blue
Green/yellow = PE
Connection of the signal lines:
The signal cables, 3 m (9.84 ft) long, 3 x 0,5 mm
signals of the condensate controller, the cooling temperature and the optional flow alarm sensor).
1 = NC
2 = MC
3 = NO
Please install the signal lines separated from the power supply lines. The contact rating for the alarm
contact is max. 250 VAC 2 A.
12
Aggressive condensate is possible.
Wear protective glasses and proper protective clothing!
False supply voltage can damage the equipment. When connecting
the equipment, please ensure that the supply voltage is identical
with the information provided on the model type plate!
For the assembly of power installations with rated voltages up to
1000 V, the requirements of VDE 0100 and relevant standards and
specifications must be observed!
The main circuit is equipped with a fuse corresponding to the
nominal current (over current protection); for electrical details see
technical data.
= L
= N
2
CSS-M/W | 1.02.00
2
(approx. AWG 17) on the back
(approx. AWG 20) are numbered from 1 – 3 (Status
www.mc-techgroup.com