Siemens SOMATOM Sensation 16 Application Manual page 93

Ct scan
Hide thumbs Also See for SOMATOM Sensation 16:
Table of Contents

Advertisement

Neck
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
w
Effective dose
nd
The 2
reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined MPR
range, linked to the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
MPRthick: ThinSliceSpi
Image thickness
Distance between images
Number of images
Contrast medium IV injection
Start delay
Flow rate
Total amount 100 ml
This protocol can also be used in case of tumor suspi-
cious of the hypopharynx and larynx. It is recommended
to acquire the images in E-phonation, and reconstruct
the images with 1 mm slice width and increment of
0.7 mm for performing MPRs. This would help differ-
entiate tumor lesions from normal pharyngeal mucosa.
ThinSliceSpi
120
150
0.75 mm
5.0 mm
12.0 mm
0.75 sec.
B30s
5.0 mm
cr-ca
12.6 mGy
Male: 1.7 mSv
Female: 1.9 mSv
45 sec.
3 ml/sec.
2
nd
reconstr.
1.0 mm
B30s
0.7 mm
sagittal
3
3
20
93