Siemens SOMATOM Sensation 16 Application Manual page 123

Ct scan
Hide thumbs Also See for SOMATOM Sensation 16:
Table of Contents

Advertisement

Abdomen
kV
Effective mAs
Slice collimation
Slice width
Feed/Rotation
Rotation time
Kernel
Increment
Image order
CTDI
w
Effective dose
nd
For the 2
reconstruction the Autoload into the
MIPthin Range on the 3D Card is activated. The images
will be automatically loaded into 3D, MIPthin, and
sagittal MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
MIPthin: AngioFast
Image thickness
Distance between images
Number of images
Contrast medium IV injection
Start delay
Flow rate
Total amount 120 ml
* CARE Bolus may be used to optimize the bolus
timing. Set the ROI for monitoring scan in the abdo-
minal aorta at the beginning of the scan range
with triggering threshold of 120 HU, or use manual
triggering.
Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images.
AngioFast
120
130
1.5 mm
5.0 mm
21.0 mm
0.5 sec.
B30f
5.0 mm
cr-ca
9.1 mGy
Male: 10.8 mSv
Female: 16.4 mSv
20 sec.*
3.0 ml/sec.
2
nd
reconstr.
2.0 mm
B20f
1.5 mm
sagittal
3
3
60
123