15.3 Checking/Setting the OCM Parameters

Fresenius Medical Care 5008 OP 5/09.06
Chapter 15: Option OCM (Online Clearance Monitoring)
4
V (urea)
Displays the V (urea).
Touch the V (urea) field for entering V (urea).
5
OCM I/O
Button for starting/stopping the OCM.
6
Plasma Na
Displays the Plasma Na.
7
Kt/V
Displays the Kt/V.
8
Clearance
Displays the clearance.
9
Time until Goal Kt/V
Displays the time remaining until achieving the dialysis dose
entered for Goal Kt/V.
10
OCM
OCM menu button
In the OCM menu
The following parameters can be checked/set.
Goal Kt/V
Interval
Hct
V (urea)
For the calculation of V (urea)
Weight
Height
Age
Sex
Note
We recommend entering V (urea). If V (urea) is not known, it will be
calculated by the system from the weight, height, age and sex using the
formula developed by Watson. Alternatively V (urea) can be determined
in the laboratory from blood samples. The PC program DC-Tool
distributed by Fresenius Medical Care is available for this purpose. The
Hct value determined from a blood sample prior to the treatment must
be entered in the Hct field.
Dialysis dose prescribed by the physician
Time between two measuring cycles
For the calculation of K and Plasma-Na
Volume of urea distribution of the patient; to be
entered in liters
For the calculation of V (urea)
For the calculation of V (urea)
For the calculation of V (urea)
For the calculation of V (urea)
+