5.2

Preparation with Rinse Solution Bag

5.2.1
Turning the Hemodialysis System On
Fresenius Medical Care 5008 OP 5/09.06
If the 5008 ONLINEplus™ hemodialysis system cannot be operated
without rinse solution bags, rinse or infusion solutions provided in NaCl
bags may be used instead.
Caution
The stability of the 5008 hemodialysis system must be ensured.
Establish the water and power supply.
Press the On/Off key. (Turn the hemodialysis system on!)
The On/Off LED is illuminated.
START-UP SCREEN
The display shows the machine type, the current software version and
the clinical data (on request) for approx. 15 seconds.
Caution
After a downtime of more than 72 hours, a cleaning program must be
performed completely before starting the treatment.
If necessary, check the hemodialysis system for presence of residual
disinfectant. (see chapter 8 Cleaning).
Note
If the message: Defective battery is acknowledged by pressing the Skip
key, it might be that the audible alarm will not be generated, if a power
failure occurs.
Chapter 5: Preparation
5-17