Compensation Factor - Kanomax TABmaster 6715 Instruction Manual

Hide thumbs Also See for TABmaster 6715:
Table of Contents

Advertisement

Note:
1) Standard Test (Std): in the condition of atmosphere 101.3kPa and temperature 21.1℃.
2) Actual Test(Act): in actual condition with the compensation for the atmosphere and temperature.
3) No Std/Act selection when in Micromanometer testing.
4) When doing actual testing under Velocity Matrix and pitot tube, manually input the current
temperature is needed for the compensation. Refer to the figures as below:
Test Setting
1.
Select Tool
2.
Test ID
3.
Test Mode
4.
Unit
5.
Data Record
6.
Std/Act
7.
K factor
8.
Temperature

6.2.7. Compensation factor

Test Setting
1.
Select Tool
2.
Test ID
3.
Test Mode
4.
Unit
5.
Data Record
6.
Std/Act
7.
K factor
Test Setting
1.
Select Tool
2.
Test ID
3.
Test Mode
4.
Unit
5.
Data Record
6.
Std/Act
7.
K factor
Note:
1) Under Capture Hood and Velocity Grid testing, K factor is used for testing data compensation in
order to get more accuracy result. And the testing range will be 0.500~1.500.
Note: K factor will affect the accuracy of measurement, so, it's necessary for having the test bench for
standard reference and using with caution.
For example:
If the Capture Hood is as the selected testing tool,
the standard airflow is 1020m
this K factor to correct the accuracy of the measurement by Model 6715.
2) Under Pitot tube testing, K factor is the factory standard coefficient and each pitot with its own
independent factor.
3) No K factor setting under the micromanometer testing.
UP
1.
Select Tool
2.
Test ID
DOWN
3.
Test Mode
RunAvg
4.
Unit
SET
5.
Data Record
OK
6.
Std/Act
Act
7.
K factor
1.000
BACK
23.5 ℃
8. Temperature
UP
DOWN
RunAvg
SET
OK
Act
1.000
BACK
UP
DOWN
RunAvg
SET
OK
Act
1.000
BACK
3
/h,at this point, K factor= standard value/testing value =1.020,input
Test Setting
UP
DOWN
RunAvg
SET
OK
Act
1.000
BACK
25.0
Test Setting
UP
1.
Select Tool
2.
Test ID
DOWN
3.
Test Mode
RunAvg
4.
Unit
SET
5.
Data Record
OK
6.
Std/Act
Act
7.
K factor
1.010
BACK
Test Setting
UP
1.
Select Tool
2.
Test ID
DOWN
3.
Test Mode
RunAvg
4.
Unit
SET
5.
Data Record
OK
6.
Std/Act
Act
7.
K factor
1.010
BACK
testing value by Model 6715 is 1000m
22
Test Setting
1.
Select Tool
2.
Test ID
3.
Test Mode
RunAvg
4.
Unit
5.
Data Record
6.
Std/Act
Act
7.
K factor
1.000
25.0 ℃
8.
Temperature
(1)Press
or
for options cha nge.
(2)Press
for save settings .
(3)Press
for quit.
UP
DOWN
SET
OK
BACK
3
/h,