Sequence to extract the sensor out of the ThermoJacket
28
info@Raytek-Direct.ca
Open the snap locks, release the rear ThermoJacket cover and leave the rear cover hang down,
supported by flexible wire rope.
Unscrew all top cable grommet compression seal nuts and push them downwards along the cables.
Use a spanner with 30 mm (1.2 in) wrench size to release the sealed cables.
Unscrew all cable connectors from the Endurance sensor by hand power and pull the cables completely
out of the ThermoJacket grommet.
Turn the Endurance sensor by hand power, induced to one of both cable sockets, some degrees anti-
clockwise to give free space for the afterwards attached extraction tool.
Press the flexible rear cover wire rope and the attached screw eye with your left-hand thumb to the left
ThermoJacket inside.
Screw the extraction tool by right hand power onto the left M16 Endurance sensor socket. Use the
whole thread pitch to fix the tool on the socket.
Keep the flexible wire rope and the screw eye in the outmost left position by your left-hand thumb.
Force to turn the Endurance sensor via the installed extraction tool in clockwise rotating direction, to
induce an ejection force over a die cast metal wedge.
Pull the released Endurance sensor easily by hand power out of the ThermoJacket.