3.3.2 Reassembly of the Tube Fittings

You may disassemble and reassemble a Swagelok tube fitting as often as required.

3.4 Air Cooling

The H
O ports are also designed to use air for cooling, for further information see section 3.3
2
page 17.
To control air flow, use the air flow regulator accessory, see section 4.9

3.5 Air Purging

Connect the air line to the female 1/4" NPT air-in port. Control the air flow with an air flow regulator. Air flow for
purging should be approximately 46 l / min (1.6 foot³ / min) and filtered through a pressure regulator with an integral
filter to prevent oil and contaminant build-up on the sensor lens. Connect the air flow regulator between the
ThermoJacket and the air pressure regulator. To ensure leak-free connections, use Teflon tape or equivalent pipe
sealant.
18
info@Raytek-Direct.ca
1.
Insert tubing with pre-swaged ferrules into the fitting
body until the front ferrule seats.
2.
Rotate the nut with a wrench to the previously pulled-up
position. At this point, a significant increase in
resistance will be encountered.
3.
Tighten slightly with a wrench.
Air Flow Regulator
Water
Cooling,
(A-TJ-AFR), page 33.