Reset Command; Setup Arm Command - Racal Instruments 1260-14C Manual

1260 vxi switching card, open collector digital i/o module
Table of Contents

Advertisement

RESET command

SETUP ARM
command
Module Specific Syntax 3-12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Syntax:
RES[ET]
Description:
The RESET command resets the 1260-14C card to
the power-up state. Specifically, the following attributes are
programmed after the RESET command is executed:
BUSY Polarity
CLKIN Polarity
Synchronous Ports
Arm
All Ports
Syntax:
SE[TUP] <address>.AR[M],ON | OFF
<address>::= Module address of the 1260-14C (1-12)
Description:
This command is used to arm and disarm the
synchronous handshake mode. Setup data may only be modified
while ARM is OFF. Synchronous data transfers may only take
place when ARM is ON. Once the card is armed, any attempts to
send a setup command other than a SETUP <address>.ARM,OFF
will cause an error.
The completion of the last synchronous READ/WRITE operation in
a test automatically sets the ARM mode to OFF. The CLKIN
signal used to cause data transfers is ignored as long as ARM is
OFF.
Each time the ARM is set to ON, synchronous READ/WRITE
operations restart at the first location in the port's buffer. This
means that if the user sets up a port to output five data items,
resetting the ARM causes the first data item to be transferred at
the next occurrence of the CLKIN signal. This is regardless of
where in the buffer the test had been when the ARM was set to
OFF. PDATAOUT commands are not recognized until ARM is
OFF. The default power-up condition of ARM is OFF.
Example:
This command sets ARM mode to ON within the module at
address 1.
SETUP 1.ARM,ON
User Manual 1260-14C
Positive
Positive
None (SYNC 0)
Off
Tri-stated

Advertisement

Table of Contents
loading