Module Specific Syntax; Syntax Notations And Conventions - Racal Instruments 1260-14C Manual

1260 vxi switching card, open collector digital i/o module
Table of Contents

Advertisement

User Manual 1260-14C
Module Specific
Syntax
Syntax Notations
and Conventions
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The 1260-14C Digital I/O module supports the PDATAOUT,
PSETUP, READ, RESET, SETUP, and WRITE commands. The
general form of the module specific syntax for the 1260-14CC
Digital I/O module is:
Command address.parameters
where:
command is one of the 1260-14C module specific commands.
Address is the module address set by SW1 on the 1260-14C
(1-12).
Parameters are the command specific parameters and data.
The address used here is not the VXIbus defined
logical address of the Master. It is a designation
unique to the 1260 Series and is used by the 1260
Option 01 to identify individual 1260 modules. This
switch setting is a requirement, since a single Option
01 can control multiple 1260 Series modules. This
address corresponds to the binary value of the switch
setting of SW1 on the switching module PCB, and
can take on the values from 1 to 12.
The following syntax notations and conventions are used
throughout the manual:
All terms in upper case letters are used directly in the
command syntax, although case is not important in the
actual command string sent to the Option 01.
All items in brackets ( [ ] ) are optional.
All lower case letters within greater/less than pairs ( < > )
are to be replaced by numeric values, as specified in the
syntax description.
A vertical bar ( | ) represents the "or" function, and specifies
that the user must select only one of the items separated
by bars.
All lines must be terminated by an ASCII line feed <LF>,
EOI or both.
NOTE

Module Specific Syntax 3-5

Advertisement