Haier HBD-116 Operation Manual page 16

Vaccine/water-pack freezer (deep freezer)
Table of Contents

Advertisement

Default set temperature
Default set temperature of this series of product is different due to difference of actual volume and
refrigerating system. Under normal condition, the best performance will be obtained if it is operated
in accordance with temperature set in the table below.
Model
Default set temperature
Enquiry of historical data
1,Enquiry of highest & lowest temperature in 24 hours:
a, In the unlock state, press "SET" for over 3 seconds, display will show code "PS";
b, Press "
" or "
" to select "Ho1" or "Lo1"( scroll display PS, AB, Po2, Ho1, Lo1......);
c, Press "SET" to confirm, the selected historical data will show up :
"Ho1"——highest temperature in 24 hours;
"Lo1" —— lowest temperature in 24 hours,
(If freezer is powered off for over 24 hours or battery runs out for 24 hours,
history data will be cleared.)
d, Press "SET" for over 3 s, it will quit setting mode.
2,Enquiry of historical alarm data:
a,In the unlock state, press both " SET" and "
b,If there is no historical alarm data, display shows "noA"; otherwise, it shows "A-X". The first
historical alarm data A-1 is the latest one and other alarm data are named by this analogy. Press
"
" or "
" to select the alarm data. Press "SET" button, the display will show the selected alarm data
and alarm time alternately.
c,Alarm time has decimal point, with integer part for hour and fractional part for 10 minutes.
For example: 3.5 means that an alarm occurs 3 hours and 50 minutes ago; 0.3 means that an alarm
occurred 30 minutes ago.
Alarm code and meaning:
Hi
Lo
14
HBD-116
-19
" for 3s, historical alarm data can be enquired.
High temperature alarm
Low temperature alarm
HBD-286
-18