Calibration procedure

NOTES

Preparing/updating the list of calibration standards

CANNON
Calibration is essential for the proper operation of the CCS. The CCS is
calibrated before it is shipped to the customer; however, the CCS should
be recalibrated after installation. Frequency of recalibration is at the
discretion of the user.
The CCS calibration procedure ensures that the mathematical constants
used in calculation of viscosity from CCS measurements accurately
reflect the unique properties of the CCS rotor/stator mechanism.
During calibration, the software uses calibration standard data from CCS
tests of viscosity standards to perform the mathematical calculations
necessary to derive three calibration constants– b0, b1 and b2. Then the
program displays the new constants and recomputes a viscosity (in cSt)
for each of the calibration standards. The resulting viscosity calculation is
compared with the known viscosity of the standard and displayed on the
computer screen.
The software allows the operator to manually change the constants if
necessary to more closely match actual and calculated viscosities.
A calibration must be performed for each temperature at which samples
will be tested.
During installation of your VISCPRO
CANNON CL standards (see table, next page) is automatically loaded
into the VISCPRO
be necessary for you to enter the viscosity data from the standard bottles
into the VISCPRO
1. Click Configure from the VISCPRO
your CCS instrument.
2. Click CCS Calibration Standards from the Configure menu
options. The CCS Calibration Standards window will appear (see
graphic, following).
3. Assemble your calibration standards and input the viscosity data (in
mPa·s) from the standard bottles for each CL standard and tempera-
ture in the appropriate fields. If you have additional standards,
including user-defined standards, you can enter these by right-
®
CCS-2100 with VISCPRO
Version 2.0a— November, 2010; CANNON
2139 High Tech Road • State College, PA 16803 • USA

CALIBRATING THE CCS

®
software package, a list of current
®
database. Before testing calibration standards, it will
®
program using the following procedure:
®
®
Instruction & Operation Manual
®
Instrument Company
primary display and select
23