Test 2-2: Dc Isolation Test - Agilent Technologies 75000 C Series Service Manual

32-channel, 5-amp, form c switch
Hide thumbs Also See for 75000 C Series:
Table of Contents

Advertisement

Test 2-2:
DC Isolation Test
NOTE
NC (&C) to NO Isolation
28 Verification Tests
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This test verifies that sufficient DC isolation exists at various points on the
Module. DC Isolation is checked from NO to NC, NO to Chassis, and NC to
Chassis. This test uses the test fixture (see Figure 2-1).
The DMM used should be capable of measuring at least 1 G
DMM indicates an overload, record the reading as >Rmax, where Rmax is
the highest resistance that the DMM can measure. For example, if the
DMM is an Agilent 3458A, a typical return for an overload is 1.E+38 and
the entry in Table 2-1 should be >1.2 G
Make hardware connections as shown in Figure 2-4
1.
Set DMM to 2-wire ohms, 1 G
2.
Send
*RST
to Module
3.
Trigger the DMM with
4.
Record the DMM reading in Table 2-1 (NC to NO)
5.
Figure 2-4. NC to NO Isolation
..
range
TRIG SGL
.. If the
Chapter 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1463a