Set DMM to measure 4-wire Ohms
Connect DMM leads as shown in Figure 2-3
Turn mainframe power ON
Measure Channel 0 C to NC Resistance
2.
Send
to Module
*RST
Send
CLOS (@nn00:nn31)
# (typically 01)
Send
OPEN (@nn00)
(typically 01)
Trigger the DMM with
Enter the result in Table 2-1 for Channel 0 C to NC
Send
CLOS (@nn00)
(typically 01)
Repeat for Channels 1 through 31
3.
Repeat step 2 for channel 1 through 31 C to NC
Use
OPEN (@nncc)
cc
= channel # (omit leading zeroes in
Figure 2-3. Common to NC Test Connections
to close all channels, where
to open channel 0, where
TRIG SGL
and note reading
to close channel 0, where
and
CLOS (@nncc)
, where
nn
)
Verification Tests 23
nn
= card
nn
= card #
nn
= card #
nn
= card # and