Digital Pulse Input Trigger

In this sampling scenario, a sampling program will automatically start when a pulse is received from a digital
device. The program will continue until it is complete, even if a pulse is only sent once from the digital device to the
ProSample. Therefore, a digital pulse essentially triggers a stand-alone sampling program to start.
This sampling scenario can only be completed with the ProSample P-12. The SDI-12 cable with flying leads (item
#630146) must be used. Please refer to the
assistance when setting up this sampling scenario.
NOTE: The settings modified in the section
and every program, so it is important these settings are carefully set before starting any program.
NOTE: Screenshots demonstrate an example setup. Settings should be adjusted for your specific application.
42
SDI-12 Cable with Flying Leads
Configure Key Instrument Settings
1. From the Main Menu, select Setup g System Settings g
Program Inputs g Program Input.
2. Highlight Prg. Start Pulse and press the Enter key.
3. Select a Program Input. If no other programs will be
configured (e.g. SDI-12 or analog programs), Program 1 Start
can be selected. Otherwise, select an appropriate Program
Input given the other programs that will be configured.
4. Press the Enter key. This will add a checkmark next to the
appropriate Program Input.
5. Go back to the Main Menu by pressing the Back key a few
times.
6. Select Programs g Change. Choose the program you would
like to modify by using the left and right arrow keys. The
program number selected must match the program number set
in steps #3 and 4.
section on page 26 or
on page 22 will be applied to each
Programming: Digital Pulse Input Trigger
contact YSI
for