Calibrating the Analog Signal

The analog signal coming to the ProSample from an analog device must be calibrated before configuring any analog
program. To calibrate the analog signal:
Wiring and Connecting Analog, Digital I/O, and SDI-12 Devices
1. From the Main Menu, select Setup
Analog Signal
Calibration
g
2. Analog 4-20 mA and Analog 0-20 mA are available options.
Highlight the appropriate option for your device and press the
Enter key.
3. On your analog device, choose to output a mA value that
corresponds to the low value of the range selected in step #3.
As an example, if 4-20 mA was selected in step #3, output a 4
mA signal from your analog device.
4. The Measured Value on the ProSample display should equal
(or nearly equal) the desired mA value. Press the Enter key to
confirm.
As an example, if 4 mA is the signal sent from the analog
device, the Measured Value might be equal to 3.9 mA. Please
note the screenshot below was taken before the mA device
sent a 4 mA signal to the ProSample.
5. On your analog device, choose to output 20 mA.
6. The Measured Value on the ProSample display should equal
(or nearly equal) 20 mA. Press the Enter key to confirm.
7. The ProSample will display Calibration Done. Press the Enter
key to return to the System Settings menu.