Stand-Alone Sampling

The ProSample will collect samples at a user-defined interval. All ProSample models can be used for this sampling
scenario. A flow signal cable and/or SDI-12 cable is not required to run this program.
NOTE: The settings modified in the section
and every program, so it is important these settings are carefully set before starting any program.
NOTE: Screenshots demonstrate an example setup. Settings should be adjusted for your specific application.
28
Configure Key Instrument Settings
1. From the Main Menu, select Program
program you would like to modify by using the left and right
arrow keys. In this example, we will modify Program # 1. Press
the Enter key to begin modifying the program.
2. Highlight Time and press the Enter key.
3. Enter a Sampling Interval. This determines how often a sample
will be collected and is determined by the user based on the
application and sampling requirements. Press the Enter key to
begin modifying the program.
4. The Sample Volume is displayed. This is the amount of sample
that will be collected at each Sampling Interval. The Sample
Volume is determined by the user based on the application and
sampling requirements. Press the Enter key to begin modifying
the program.
on page 22 will be applied to each
Change. Choose the
g
Programming: Stand-Alone Sampling