Table 4: Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The
ConMed Linvatec
electromagnetic environment specified below. The customer or the user of this
VP47XXF
Monitor should assure that it is used in such an environment.
IEC 60601
Immunity Test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power
transients / bursts
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV line to line
IEC 61000-4-5
± 2 kV lines to earth
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage dips, short
<5%
interruptions and
(>95% dip in
voltage variations
for 0.5 cycle
on power supply
40%
input lines
(60% dip in
IEC 61000-4-11
for 5 cycles
70%
(30% dip in
for 25 cycles
<5%
(>95% dip in
for 5 seconds
Ut
Note:
is the a.c. mains voltage prior to application of the test level.
VP47XX
and
VP47XXF
Compliance Level
Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line to line
± 2 kV lines to earth
3 A/m
Ut
Ut
<5%
Ut
)
(>95% dip in
for 0.5 cycle
Ut
Ut
40%
Ut
)
(60% dip in
for 5 cycles
Ut
Ut
70%
Ut
)
(30% dip in
for 25 cycles
Ut
Ut
<5%
Ut
)
(>95% dip in
for 5 seconds
E-43
Monitors are intended for use in the
Electromagnetic environment
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
Ut
)
environment. If the user of the
VP47XX
and
requires continued operation
Ut
)
during power mains interruptions,
it is recommended that
and
VP47XXF
powered from an uninterruptable
Ut
)
power supply or battery.
Ut
)
VP47XX /
VP47XXF
Monitors
VP47XX
Monitors be
E

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Linvatec VP4726F and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Linvatec vp4732Linvatec vp4732fLinvatec vp4726

Table of Contents