QBC STAR™ System User's Manual

Disable Auto LCD (Contrast)

As the QBC STAR instrument is used, it may warm up several degrees. This can cause the LCD screen
contrast to change. The instrument monitors operating temperatures and automatically adjusts the
contrast. The Disable Auto LCD (Contrast) option lets you enable (turn on) or disable (turn off) the
automatic contrast adjustment of the LCD screen.
When you select the option, a message tells you whether the auto contrast is currently enabled or
disabled. Press the  or  key to change the status from enabled to disabled or vice versa. When the
desired status is shown, press

Save Data

This option is for factory use only.

Print Data

This option is typically used to help QBC Diagnostics Inc. personnel to troubleshoot possible problems.
When selected, "Print Data" sends raw scan data from tube readings and diagnostic parameters to the
selected printer. The information is useful only to QBC Diagnostics Inc. technical personnel in trying to
determine the source of errors that may occur.

Shipping Prep.

This option prepares the instrument to be physically shipped. It performs three functions. First, it al-
lows the rotor to spin freely. Second, it moves the optics carriage back so that protective foam can be
inserted. Third, it unlocks the door for the insertion of the foam.
After you select this option, a message advises you to turn off instrument power prior to inserting the
foam.

More Options

If you select number 7 – More Options, a new menu appears with additional selections. See the sample
graphic below.