Download Print this page

Emerson Rosemount 3051 Quick Start Manual page 14

Pressure transmitter and dp flow meters with wirelesshart protocol
Hide thumbs Also See for Rosemount 3051:

Advertisement

Quick Start Guide
2.5.2 Verify transmitter configuration using Field Communicator
For HART Wireless transmitter communication, a Rosemount 3051 Wireless
device descriptor is required. For connecting with a Field Communicator, refer
to
Figure
1-1.
From the Home screen, enter the Fast Key sequence:
Device dashboard Fast Keys
Table 2-2: Device Revision 1, DD Revision 1 Fast Keys
Function
Tag
Date
Descriptor
Message
Long Tag
Network ID
Join Device to Network
Update Rate
Range Values
Transfer Function
Units
Lower Sensor Trim
Upper Sensor Trim
Digital Zero Trim
Rerange by Applied Pressure
Custom Display Configuration
Scaled Variable
Find Device
Simulate Digital Signal
2.5.3 Verify transmitter configuration using Emerson Wireless Gateway
In the Gateway's integrated web interface, navigate to the Explorer > Status
page. This page will show whether the device has joined the network and if it
is communicating properly.
14
February 2019
3, 5
Fast Keys
2, 1, 1, 1, 1
2, 1, 1, 1, 5
2, 1, 1, 1, 3
2, 1, 1, 1, 4
2, 1, 1, 1, 2
2, 2, 1, 1
2, 2, 1, 2
2, 1, 4
2, 1, 1, 5
2, 1, 1,6
2, 1, 1, 2
3, 5, 1, 1, 2
3, 5, 1, 1, 1
3, 5, 1, 1, 3
2, 2, 2, 2, 1
2, 1, 5
2, 1, 7, 1
3, 5, 2
3, 6
Emerson.com/Rosemount

Advertisement

loading

This manual is also suitable for:

Rosemount 3051cf