7. Within the Calibrate tab, click on the
Defect tab, then click the ReadStatus
button.
8. Highlight Index 1, then click the
SelectFile button. This will change the
Index 1 status to enable.
9. Click the DownloadFile button, and
then click the Path button from the
popup window.
Customer Support
[ViZion DR+ 1417V2 Wireless Panel Installation Manual]
10. Browse to the
C:\opal\data\KGxxxxxxxxxxx\Correct
\Default folder and select the
defect_2304x2800.gn file. Click Open.
Note: if the defect file does not appear,
verify the file type dropdown is set
to defect files(*.dft).
11. Type 1 in the FileIndex field and then
click OK.
12. After the defect file has been
downloaded to the panel. Click OK.