4.7.4 Panel Calibration Setup
1. Select the Acquire tab in iDetector.
2. Uncheck the Defect, Gain, and Offset
checkboxes
in that
Note: during the calibration, please do not
move the detector once calibration
has been started and keep Link
cable connected to the detector.
[ViZion DR+ 1417V2 Wireless Panel Installation Manual]
order.
Customer Support
3. Select the Calibrate tab, check the
radio button for PostOffset, and click
the Start Generate Templates button.
The Calibration wizard will open into a new
window, presenting the Welcome screen.
4. Set X-ray environment:
Tube SID = 50" or higher (at least
greater than 47").
No Collimation = open collimator
blades at least 1" past the edges of
the panel on all sides.
No Grid = remove any grid or
pressure grid cover.
No Objects = remove any objects
from the surface of the panel.
5. Click the arrow in the top right corner
labeled Start the journey of
Calibration.