Shutting Down And Turning On The System - AB Sciex 4000 QTRAP Hardware Manual

Table of Contents

Advertisement

Figure 1-3 Ion Optics

Shutting Down and Turning on the System

Use the following procedures if you need to shut down or turn on the system.
To shut down the system
Document Number: D1000092200 B
Curtain Gas™ flow
Q0
ST
Q1
Skimmer
Orifice
IQ1
Curtain Plate
1. In the Analyst
software, complete or stop any ongoing scans.
®
2. Turn off the sample flow to the instrument.
CAUTION! Potential Instrument Damage: Shut off the sample flow before you
shut down the instrument.
3. In the Analyst software, deactivate the hardware profile and then close the
application software.
4. Stop the Analyst
®
Analyst software.)
CAUTION! Leave the roughing pump running for a minimum of 15 minutes
after turning off the instrument's main power switch. When the main power
switch is turned off, the turbo pumps continue to rotate without power for a
few minutes and continue to provide vacuum to the vacuum chamber. If, during
this time, the roughing pump is turned off, the pressure in the vacuum line
between the roughing pump and the turbo pumps increases. The increase in
back pressure can create an incorrect load on the turbo pumps' bearings and
can cause a catastrophic failure of the turbo pumps.
CAUTION! If the instrument is to be shut down for any length of time, we
recommend that the vacuum chamber be vented to prevent exhaust from the
roughing pump being sucked back into the vacuum chamber. To vent the
vacuum chamber, follow steps 5 to 7.
Q2
LINAC®
Collision
ST2
Cell
ST3
Q3
IQ2
IQ3
Aux RF
Service. (See the Software Reference Guide for the
Hardware Guide
Exit
CEM
Lens
Deflector
9

Advertisement

Table of Contents
loading

Table of Contents