Models V350, V600 & V800: Other Procedures

Emergency Stop Switch

A red emergency stop switch is located on the inside of the mast.
WARNING
If the hand control fails and the lift continues to raise or lower, push down on the
emergency stop switch to stop the lift.

Base Adjustment

The base of the VERA-LIFT™ should be open to its widest position for all transfers. The base
may be narrowed to go through doorways but must be opened after the lift has cleared the
door. To adjust the width of the base, stand behind the lift and grasp the soft rubber handle
on the spreader bar located at the rear of the lift. Adjust as follows:
Closed
Half Open
Open

Rear Caster Brakes

The VERA-LIFT™ is equipped with caster brakes on the rear wheels. To lock the brakes,
step down on the end of each brake tab. To release the brakes, push forward on the tab
located on the top of each brake.
WARNING
The brakes need to be locked as the patient is being lifted and should remain unlocked
as the patient is being lowered.

Emergency Up/Down Switch

WARNING
If the pendant switch fails, use the emergency up/down switch located on the back of
the lift base. To use it, push up on the switch to raise the lift and down to lower the lift.
This switch should only be used in emergency situations when the pendant switch fails.
Do not press on the hand held pendant switch buttons and on the
emergency up/down switches at the same time.
April 2013
For your nearest Distributor, call 1-800-694-4525
November 2017
Move the spreader bar to the right hand position.
Move the spreader bar to the center position.
Move the spreader bar to the left hand position.
For your nearest distributor, call 1-800-694-4525