Emergency stop

To activate emergency stop:
Depress the red emergency stop button on the control box.
Resetting:
Turn the button in the direction of the arrows until the button pops out.

Emergency lowering

Manual emergency lowering:
For manual emergency, turn the round plastic knob on the actuator clockwise.
Electrical emergency lowering:
For electrical emergency, use the down button on the control box.
Trouble-shooting
If the lift or base-width adjustment cannot be activated, check the following:
- That the emergency stop button is not pressed in.
- That all cables are properly and securely connected. Pull out the contact and plug it in again firmly.
- That battery charging is not in progress.
- That the battery is charged.
If the lift is not working properly, contact your dealer.
If the lift makes unusual noises:
- Try to determine the source of the sound. Take the lift out of operation and contact your dealer.