6)
Lower the lift arms to their lowest position and spread the base of the VERA-LIFT™
to its widest position. (Spreading the base maximizes the stability of the lift.)
7) Move the lift in front of the patient and help the patient place his or her feet on the foot
pad with the patient's knees against the knee pad.
It is important that the patient's feet be flat on the base of the foot pad.
(The height of the foot pad can be raised by positioning it on the upper mast pegs for
VERA-LIFT™ models without scales.)
8)
Lock the caster brakes.
9)
Connect the back belt to the lift arm hooks snuggly:
a) Cross one back belt white strap through the loop closest to the back belt on the
other white strap.
b) Connect the right side white back belt loops to the left lift arm hook and connect the
left side white back belt loops to the right lift arm hook using the same loop on both
sides of the back belt.
c) Remove the colored straps from the VELCRO pocket under the patient's arms and
connect the colored straps straight (not crossed) to the lift arm hooks using the
same loop on both sides of the back belt. (The upper or lower lift arm hooks may be
used for short to medium height patients. The upper hooks should be used for tall
patients as this will help the patient to stand erect on the VERA-LIFT™.)
10)
Snuggly fasten the knee strap behind the patient's knees.
Double check the back belt loop connection to the lift arm hooks to make
11)
sure the loops are connected securely to the lift.
12) Lift the patient to a standing position.
13) Unlock the VERA-LIFT™ brakes and back the lift away.
14)
While slowly turning the lift, move the patient to the desired location within
the room with the back of the patient's legs against the bed, wheelchair, etc.
15) Leaving the caster brakes unlocked, lower the patient so that his or her buttocks are
positioned correctly on the bed or at the back of the wheelchair seat, etc.
16) Remove the back belt from the lift arm hooks, unfasten the knee strap and back the lift
away from the patient.
17) Unfasten the inner safety belt and remove the back belt.
18) Make sure the patient is safely positioned before removing the VERA-LIFT™ and back
belt from the room.