Care Fusion Alaris GW Directions For Use Manual page 9

Volumetric pump
Table of Contents

Advertisement

Earth Conductor
The Alaris® GW Volumetric Pump is a Class I device, therefore must be earthed when connected to an AC
d
power supply.
This pump also has an internal power source.
When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If the
integrity of the external protective conductor on the AC power cable has been compromised, the pump
should be disconnected from the AC power source and operated utilising the internal battery.
Hazards
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise care to
locate the pump away from any such hazardous sources.
A fire hazard may exist if the pump is used in the presence of high oxygen concentrations.
Dangerous Voltage: An electrical shock hazard exists if the pump's casing is opened or removed. Refer all
m
servicing to qualified service personnel.
Electrostatic discharge (ESD) precautions are required when connecting RS232/Nurse Call. Touching the pins
V
of the connectors may result in ESD protection failure. It is recommended that all actions must be taken by
appropriately qualified service personnel.
If this pump is dropped, subjected to excessive moisture, fluid spillage, humidity or high temperature, or
*
otherwise suspected to have been damaged, remove it from service for inspection by qualified service
personnel. When transporting or storing the pump, use original packaging where possible, and adhere to
temperature, humidity and pressure ranges stated in the Specifications section and on the outer packaging.
Warning: Alaris® GW Volumetric Pumps should not be modified or altered in any way, except where explicitly
directed or authorised by CareFusion. Any use of Alaris® GW Volumetric Pumps which have been altered or
modified otherwise than in strict application of directions provided by CareFusion, is at your sole risk, and
CareFusion does not provide any warranty for or endorsement on any Alaris® GW Volumetric Pump that
has been so modified or altered. CareFusion's product warranty shall not apply in the event the Alaris® GW
Volumetric Pump has suffered damage or premature wear, or malfunctions or otherwise operates incorrectly,
as a result of unauthorised modification or alteration of the Alaris® GW Volumetric Pump.
1000DF00430 Issue 4
Alaris® GW Volumetric Pumps
Operating Precautions
8/36