Starting the Infusion without a Flow Sensor

Check:
• The pump is switched on.
• The Infusion set has been primed (refer to 'Priming the Infusion Set' section of this DFU).
• The in-line clamp is open.
indicates infusion without a flow sensor being used.
Standard Mode
1. Enter infusion rate using the
2. Press
button once to confirm the infusion rate.
e
3. Enter VTBI using the
f h
4. Press
button to confirm the VTBI.
e
5. Press
to clear VI if required.
c
6. Press
button to start infusing.
d
Standard Mode with VTBI / Time Infusion on
1. Enter VTBI using the
f h
2. Press
button once to confirm the VTBI.
e
3. Enter TIME using the
f h
4. Press
button to confirm the TIME.
e
5. Press
to clear VI if required.
c
6. Press
button to start infusing.
d
Micro Mode
1. Enter infusion rate using the
2. Press
button once to confirm the infusion rate.
e
3. Enter VTBI using the
f h
4. Press
button to confirm the VTBI.
e
5. Press
to clear VI if required.
c
6. Press
button to start infusing.
d
Micro Mode with VTBI / Time Infusion on
1. Enter VTBI using the
f h
2. Press
button once to confirm the VTBI.
e
3. Enter TIME using the
f h
4. Press
button to confirm the TIME.
e
5. Press
to clear VI if required.
c
6. Press
button to start infusing.
d
keys.
f h
keys.
keys.
keys.
keys.
f h
keys.
keys.
keys.
1000DF00430 Issue 4
Alaris® GW Volumetric Pumps
Starting the Infusion without a Flow Sensor
13/36