Flow Sensor Operation

The flow sensor automatically monitors the infusion flow rate through the drip chamber. The flow sensor will cause
A
the pump to alarm if a significant deviation from the set infusion rate occurs. The flow sensor will also be able to
detect empty containers. For this reason, use of a flow sensor is recommended when using an infusion set without
an anti-siphon valve.
Flow Sensor Usage
Infusion set with anti-siphon valve ?
YES
NO
Model 180 Flow Sensor
Pull back handles
1. Plug the flow sensor into the flow sensor interface located on the top rear part of the pump.
2. Attach the Model 180 Flow Sensor to the drip chamber of the Infusion set, by pulling back the handles. Refer to the illustration
above.
3. Proceed with load, priming, and set-up instructions as described in section 'Getting Started' .
NOTE: Ensure drip chamber is half full and upright.
Always attach the flow sensor before you start an infusion.
A
Avoid using the flow sensor in direct sunlight.
Always ensure lens is clean.
Always replace the flow sensor interface cover when the flow sensor is disconnected.
Use flow sensor ?
Optional
Recommended
Handles
Flow sensor
interface
1000DF00430 Issue 4
Flanges
Drip chamber
Flow
Sensor
Interface
Cover
22/36
Alaris® GW Volumetric Pumps
Flow Sensor Operation