User Selectable Options - Care Fusion Alaris GW Directions For Use Manual

Volumetric pump
Table of Contents

Advertisement

User Selectable Options

To set the user options the pump must be on
A
enabled. See also 'Configurable Options' section in this DFU.
Press
button and hold for 2 seconds to enter User Selectable Options.
e
Note: Number of
button presses are dependent upon the User Select Mode Options which have been enabled. The following
e
instructions are based upon all options being enabled.
Setting the Occlusion Pressure Level
1. When PrES is displayed.
2. Select HI, nor or Lo using the
3. Press e button to return to
Setting the Alarm Volume
button until tonE is displayed.
1. Press
e
2. Select an alarm volume between 1 (Low) and 7 (High) using the
3. Press
button to return to
e
Setting a VTBI / Time Infusion
1. Press
e
button until
R
2. Select On or OFF using the
3. Press
button to return to
e
4. Make sure that the TIME indicator is lit if set to on.
Note: The TIME increases / decreases in rate dependant units e.g. 10ml @ 99.9ml/h is 6mins, therefore 0:06 is displayed.
Setting to Micro Mode
1. Press
button until
e
T
2. Select On or OFF using the
3. Press
button to return to
e
4. Make sure that the MICRO indicator is lit if set to on.
keys to set the occlusion pressure level to High, Normal or Low.
h f
or set-up mode or next option.
G
or set-up mode, or next option.
G
is displayed.
keys to turn the VTBI / time infusion setting on or off.
h f
or set-up mode, or next option.
G
is displayed.
keys to turn micro mode on or off.
h f
or set-up mode, an audible sounder will confirm the status.
G
1000DF00430 Issue 4
or in set-up mode, and the relevant user selectable options are
G
keys.
h f
18/36
Alaris® GW Volumetric Pumps
User Selectable Options