Warnings - Care Fusion Alaris GW Directions For Use Manual

Volumetric pump
Table of Contents

Advertisement

Warnings

Warnings alert the user but may not stop the infusion and are indicated by an audible alarm, a message on the display or both.
1. Check the display for a warning message. Press
2. Rectify the cause of the warning or proceed with caution.
Display
C
D
E
F
G
H
I
to silence the alarm.
c
Cause
Bolus is being administered.
Finished pre-set volume to be infused.
The pump is priming the Infusion set.
Low Battery
(At least 30 mins before bAt alarm).
The pump is on hold.
The pump has been left unattended for 2mins
and infusion has not started.
Automatic set check.
1000DF00430 Issue 4
Alaris® GW Volumetric Pumps
Action
Release
button to return to infusion once correct
b
bolus has been administered.
Pump will infuse at the keep vein open rate, until the
button is pressed. Refer to 'KVO Rate' section of
d
this DFU.
Ensure all air has been primed out of the infusion set,
before starting the infusion.
Connect pump to an AC power source.
Press
to return to infusion, or press
d
to set-up.
Attend to pump.
Allow test to complete before operating the pump
further.
21/36
Warnings
to return