Care Fusion Alaris GW Directions For Use Manual

Volumetric pump

Advertisement

Quick Links

Alaris® GW Volumetric Pump
Directions For Use
en
s

Advertisement

Table of Contents
loading

Summary of Contents for Care Fusion Alaris GW

  • Page 1 Alaris® GW Volumetric Pump Directions For Use...
  • Page 2: Table Of Contents

    Alaris® GW Volumetric Pumps Contents Page Introduction ....................2 About This Manual .
  • Page 3: Introduction

    Alaris® GW Volumetric Pumps Introduction Introduction The Alaris® GW Volumetric Pump (herein after referred to as 'Pump') is a small lightweight volumetric infusion pump that provides accurate and reliable infusions over a range of rates. The ideal pump for general care and critical care. Intended Purpose The Alaris®...
  • Page 4: Features Of The AlarisĀ® Gw Volumetric Pump

    Alaris® GW Volumetric Pumps Features of the Alaris® GW Volumetric Pump Features of the Alaris® GW Volumetric Pump Main Display - Display Indicators Displays the infusion rate, VTBI, VI and time remaining for VTBI / Time infusions. Display will flash when pump is operating on battery.
  • Page 5: Controls And Indicators

    Alaris® GW Volumetric Pumps Controls and Indicators Controls and Indicators Controls Symbol Description ON/OFF button - Press once to switch the pump ON. Press and hold down for approximately 3 seconds to switch the pump OFF. RUN/HOLD button - Press to start the infusion or to put the infusion on hold. CLEAR/SILENCE button - Press to silence alarm for 1 minute.
  • Page 6: Symbol Definitions

    Alaris® GW Volumetric Pumps Symbol Definitions Symbol Definitions Labelling Symbols: Symbol Description Attention Consult accompanying documents Potential Equalisation (PE) Connector > RS232/Nursecall Connector. Type CF applied part. (Degree of protection against electrical shock) Protected against vertically falling drops of water Alternating Current Device complies with the requirements of Council Directive 93/42/EEC as amended by 2007/47/EC.
  • Page 7: Operating Precautions

    Alaris® GW Volumetric Pumps Operating Precautions Operating Precautions Infusion Sets • To ensure correct and accurate operation, only use CareFusion single use infusion sets described in the 'Compatible Infusion Sets' section of this Directions for Use. Use an infusion set with an anti-siphon valve whenever possible.
  • Page 8 Alaris® GW Volumetric Pumps Operating Precautions Operating Environment • When using any infusion pump in conjunction with other pumps or devices requiring vascular access, extra care is advised. Adverse delivery of medication or fluids can be caused by the substantial variation in pressures created within the local vascular system by such pumps.
  • Page 9 Alaris® GW Volumetric Pumps Operating Precautions Earth Conductor • The Alaris® GW Volumetric Pump is a Class I device, therefore must be earthed when connected to an AC power supply. • This pump also has an internal power source. • When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used.
  • Page 10: Getting Started

    Alaris® GW Volumetric Pumps Getting Started Getting Started Before operating the pump read this Directions For Use (DFU) manual carefully. Initial Set Up 1. Check that the pump is complete, undamaged and that the voltage rating specified on the label is compatible with your AC power supply.
  • Page 11 Alaris® GW Volumetric Pumps Getting Started Loading an Infusion Set Ensure the appropriate Infusion set for the fluid/drug to be infused has been selected. Follow the instructions supplied with the individual Infusion set. Use of non-specified Infusion sets may impair the operation of the pump and the accuracy of the infusion. For Alaris®...
  • Page 12 Alaris® GW Volumetric Pumps Getting Started Power On/Off To power up the pump: 1. Press the button once and release. 2. Check: • A high pitch sounder is activated for 3 seconds during this time the main speaker 'beeps' once upon power up sequence. • All display segments and all indicators are lit.
  • Page 13: Starting The Infusion With A Flow Sensor (Recommended)

    Alaris® GW Volumetric Pumps Starting the Infusion with a Flow Sensor (Recommended) Starting the Infusion with a Flow Sensor (Recommended) The flow sensor automatically monitors the infusion flow rate through the drip chamber. The flow sensor will cause the pump to alarm if a significant deviation from the set infusion rate occurs. The flow sensor will also be able to detect empty containers.
  • Page 14: Starting The Infusion Without A Flow Sensor

    Alaris® GW Volumetric Pumps Starting the Infusion without a Flow Sensor Starting the Infusion without a Flow Sensor Check: • The pump is switched on. • The Infusion set has been primed (refer to 'Priming the Infusion Set' section of this DFU). • The in-line clamp is open.
  • Page 15: Secondary / Piggyback Infusions

    Alaris® GW Volumetric Pumps Secondary / Piggyback Infusions Secondary / Piggyback Infusions Secondary (or piggyback) Infusion mode is only available if configured, refer to Configurable Options section of this DFU. Secondary Infusion mode is used to administer an intermittent fluid / drug solution e.g. 4 hourly antibiotic infusion using: • A primary infusion set with an in-line check valve before the Y-Injection site e.g.
  • Page 16: Basic Features

    Alaris® GW Volumetric Pumps Basic Features Basic Features Rate Titration 1. Enter the new infusion rate using the keys. 2. Press button to confirm the infusion rate. Note: The rate can be increased or decreased without interrupting the infusion. If the new rate selected is not confirmed the pump will revert to the current rate and no change in the infusion rate will occur.
  • Page 17 Alaris® GW Volumetric Pumps Basic Features Changing the Infusion Set 1. Press to put the pump ON HOLD. 2. Close in-line clamp and ensure the IV access to the patient is isolated. 3. Disconnect the Infusion set from the patient. 4.
  • Page 18 Alaris® GW Volumetric Pumps Basic Features Clearing Air-In-Line 1. Press button to silence the air-in-line alarm and put the pump on hold. 2. Close the in-line clamp. 3. Open the door to view the air bubble. NOTE: Air-in-line alarms can be activated by both single bubbles and bubbles accumulated over time. 4.
  • Page 19: User Selectable Options

    Alaris® GW Volumetric Pumps User Selectable Options User Selectable Options To set the user options the pump must be on or in set-up mode, and the relevant user selectable options are enabled. See also 'Configurable Options' section in this DFU. Press button and hold for 2 seconds to enter User Selectable Options.
  • Page 20: Configurable Options

    Alaris® GW Volumetric Pumps Configurable Options Configurable Options The default settings are configurable as displayed in brackets in the table below. Each of the configurable options has a code which must only be altered by qualified service personnel with reference to the technical service manual (TSM) for this product (Technical Service Manual reference: 1000SM00006).
  • Page 21: Alarms

    Alaris® GW Volumetric Pumps Alarms Alarms Alarms stop the infusion and are indicated by a combination of an audible alarm and a message on the display. 1. Check the display for an alarm message and review table below for cause and action. Press to silence the alarm.
  • Page 22: Warnings

    Alaris® GW Volumetric Pumps Warnings Warnings Warnings alert the user but may not stop the infusion and are indicated by an audible alarm, a message on the display or both. 1. Check the display for a warning message. Press to silence the alarm. 2.
  • Page 23: Flow Sensor Operation

    Alaris® GW Volumetric Pumps Flow Sensor Operation Flow Sensor Operation The flow sensor automatically monitors the infusion flow rate through the drip chamber. The flow sensor will cause the pump to alarm if a significant deviation from the set infusion rate occurs. The flow sensor will also be able to detect empty containers.
  • Page 24: Compatible Infusion Sets

    Alaris® GW Volumetric Pumps Compatible Infusion Sets Compatible Infusion Sets The pump uses standard, single-use, disposable Infusion sets with Luer-lock connectors. The user is responsible for verifying the suitability of a product used, if it is not recommended by CareFusion. CareFusion recommends the use of infusion sets with anti-siphon valves whenever possible.
  • Page 25 Alaris® GW Volumetric Pumps Compatible Infusion Sets Filter Sets 273-009 1.2µm filter set with anti-siphon valve, with 15µm filter in drip chamber. (230 cm) Burette Sets 273-103 Burette set with 1 Y site and anti-siphon valve. (220 cm) Opaque Sets 273-011 Opaque PVC infusion set with anti-siphon valve and pump segment with 15µm filter in drip chamber.
  • Page 26 Alaris® GW Volumetric Pumps Compatible Infusion Sets Filter Extension Sets C20128 Extension set with 1.2µm filter and one Y site. Rotating male Luer lock. (approx. 51 cm) C20350 Extension set with 0.2µm filter and one Y site. Rotating male Luer lock (approx. 51 cm) Low Sorbing (Polyethylene Lined) 20128E-0006 Extension set with 1.2µm filter and one SmartSite®...
  • Page 27: Associated Products

    Alaris® GW Volumetric Pumps Associated Products Associated Products The Alaris® Gateway Workstation Product SKU 80203UNS0y-xx Supply Voltage 115-230VAC, ~50-60Hz Electrical Rating 460VA (Maximum) Protection Against Electrical Shock Class 1 Classification Continuous Operation Supply to Pump 115-230V, ~50-60Hz, 60VA The Alaris® DS Docking Station Product SKU 80283UNS00-xx Supply Voltage...
  • Page 28: Maintenance

    Alaris® GW Volumetric Pumps Maintenance Maintenance Routine Maintenance Procedures To ensure that this pump remains in good operating condition, it is important to keep it clean and carry out the routine maintenance procedures described below. All servicing should only be performed by qualified service personnel with reference to the Technical Service Manual (TSM).
  • Page 29: Cleaning And Storage

    Alaris® GW Volumetric Pumps Cleaning and Storage Cleaning and Storage Cleaning the pump Before the transfer of the pump to a new patient and periodically during the use, clean the pump by wiping over with a lint-free cloth lightly dampened with warm water and a standard disinfectant / detergent solution. Do not use the following disinfectant types: • Disinfectants which are known to be corrosive to metals must not be used, these include: • NaDcc (such as Presept),...
  • Page 30: Specifications

    Alaris® GW Volumetric Pumps Specifications Specifications Electrical/Mechanical Safety Complies with IEC/EN60601-1 and IEC/EN60601-2-24. Electro Magnetic Compatibility (EMC) Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24. Electrical Safety IEC/EN 60601-1 - Typical earth leakage current 40µA. Potential Equalisation Conductor The function of the Potential Equalisation Connector (Conductor) is to provide a direct connection between the Pump and the potential equalisation busbar of the electrical installation.
  • Page 31 Alaris® GW Volumetric Pumps Specifications Battery Specifications Rechargeable NiMH (Nickel Metal Hydride). Automatically charges when the pump is connected to AC power. Battery Life - 10 hours @ 25 ml/h Battery Charging - 95% charge - < 24 hours (all conditions). Alarm Conditions SYSTEM ERROR UPSTREAM OCCLUSION...
  • Page 32 Alaris® GW Volumetric Pumps Specifications Notes: 1. All accuracy specifications are with a 95% confidence interval / 95% population, unless stated otherwise. 2. Nominal conditions are defined as: • Set Rate: 125 ml/h (25 ml/h for rate accuracy); • Disposable Type: 273-001; • Needle: 18 gauge x 40 mm;...
  • Page 33: Irda, Rs232 And Nurse Call Specification

    Alaris® GW Volumetric Pumps IrDA, RS232 and Nurse Call Specification IrDA, RS232 and Nurse Call Specification RS232 / IrDA Feature The RS232 / IrDA feature is a standard feature on Alaris® GW Volumetric Pump. It allows the pump to be monitored remotely via a suitable central monitoring or computer system.
  • Page 34: Trumpet And Flow Rate Curves

    Alaris® GW Volumetric Pumps Trumpet and Flow Rate Curves Trumpet and Flow Rate Curves In this pump, as with all infusion systems, the action of the pumping mechanism and variations cause short-term fluctuations in rate accuracy. The following curves show typical performance of the system in two ways: 1) the accuracy of fluid delivery over various time periods is measured (trumpet curves), and 2) the delay in onset of fluid flow when infusion commences (start-up curves).
  • Page 35: Technical Description

    Alaris® GW Volumetric Pumps Technical Description Technical Description The following details outline the basic safety checks designed into the pump to minimise the possibility of under or over infusions. Power on Self-Tests The pump is single fault tolerant, which means the pump will either stop and alarm, or continue to infuse uninterrupted as a result of a single fault failure.
  • Page 36: Products And Spare Parts

    Alaris® GW Volumetric Pumps Products and Spare Parts Products and Spare Parts Alaris® Infusion System Range of products in the Alaris® Infusion System product family are: Part Number Description 8002MED01 Alaris® GH Syringe Pump (with Plus Software) 8003MED01 Alaris® CC Syringe Pump (with Plus Software) 80043UN01 Alaris®...
  • Page 37: Service Contacts

    Alaris® GW Volumetric Pumps Service Contacts Service Contacts For service contact your local Affiliate Office or Distributor. CareFusion, CareFusion, CareFusion, CareFusion, PO Box 5527, Pascalstr. 2, The Crescent, Jays Close, 14B George Bourke Drive, Dubai, United Arab Emirates. 52499 Baesweiler, Basingstoke, Mt Wellington 1060, Deutschland.
  • Page 38 Alaris, Guardrails and SmartSite are registered trademarks of CareFusion Corporation or one of its subsidiaries. All rights reserved. All other trademarks are property of their respective owners. © 2000-2014 CareFusion Corporation or one of its subsidiaries. All rights reserved. This document contains proprietary information of CareFusion Corporation or one of its subsidiaries, and its receipt or possession does not convey any rights to reproduce its contents, or to manufacture...