sorin Paradym VR 8250 Implant Manual

Implantable cardioverter defibrillator

Advertisement

Implant manual
8250 model
Implantable cardioverter defibrillator

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Paradym VR 8250 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for sorin Paradym VR 8250

  • Page 1 Implant manual 8250 model Implantable cardioverter defibrillator...
  • Page 2: General Description

    3.1. POTENTIAL COMPLICATIONS of pacing to suit the patient’s activity. Complications may arise with any implanted pacing or defibrillation PARADYM VR 8250 provides a range of therapeutic and diagnostic system. They may be related to the device itself: functions: ― early battery depletion, ―...
  • Page 3 Cellular phones: Radiofrequency signals can interfere with the over the lead. The defibrillating paddles should preferably be placed functioning of PARADYM VR 8250 if the handset is placed too close in an anteroposterior position. Avoid any direct contact between the to the defibrillator.
  • Page 4: Special Modes

    One copy should be 3. Tighten and check the tightness. returned to Sorin Group within 30 days after implantation to activate Caution: 1. Do not tighten the pre-inserted screws when there is no the warranty.
  • Page 5: Main Functions

    In safety mode, the defibrillator operates with the parameters underlined in the table of programmable parameters. 6.3. RESPONSE IN THE PRESENCE OF INTERFERENCE 6.2. MAGNET MODE If the defibrillator senses electrical noise at a frequency above 16 Hz, When the magnet is applied: it switches to an asynchronous mode at the basic rate.
  • Page 6: Patient Follow-Up

    *Medical-grade materials that have undergone "in vitro" and "in vivo" qualifications. The device can be non-invasively identified as follows: 1. Take an x-ray to identify the name of the manufacturer printed on the device (S = Sorin ; D = Defibrillator ; A = PARADYM). ENGLISH – – ENGLISH...
  • Page 7: Electrical Characteristics

    Stored energy (J) 10. ELECTRICAL CHARACTERISTICS V1 (Volt) V2 (Volt) Ventricular input impedance 80 kilohms ± 30 % Delivered E: Phase 1 (J) 0.29 13.9 23.8 29.3 D.C. capacitance 180 µF ± 8 % Delivered E: Phase 2 (J) 0.09 Reformation No reformation required Tolerances are 12% for voltage (25% at 0.5 J) and 30% for energy.
  • Page 8 Pacing/Sensing Values Audible alarms Values Ventricular sensitivity (mV) From 0.4 to 4 by steps of 0.2 ; 0.4 Battery depletion ON-OFF (± 50 %) High battery drain ON-OFF Ventricular amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 Abnormal lead impedance ON-OFF (± 20 %) Abnormal RV/SVC continuity ON-OFF Ventricular pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1...
  • Page 9 ♦ 11.3.1. Therapy parameters in slow VT zone Shock program Values ATP 1 program Values Shock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5- 4-5-6-7-8-9-10-12-14-16-18-20-22- ATP program OFF-Burst-Burst+Scan-Ramp- 24-26-28-30-32-34-42 (± 30 %) Ramp+Scan Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5- Number of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 4-5-6-7-8-9-10-12-14-16-18-20-22- Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 24-26-28-30-32-34-42 (±...
  • Page 10: Non Programmable Parameters

    1. The PARADYM implantable cardioverter defibrillator is only guaranteed for one implantation. Waveform Constant tilt (50% - 50%) 2. The EURID/IAPM implant form must be sent to SORIN within 30 Stored energy for the shock MAX 42 J (± 16 %) days after implantation.
  • Page 11 Via Crescentino s.n. – 13040 Saluggia (VC) – Italy Tel. : +39 161 487323 – 487095 / Fax +39 161 487524 0459 – 2007 www.sorin-crm.com...