Connectivity Settings - Abbott Afinion 2 User Manual

Table of Contents

Advertisement

Configuration

Connectivity settings

See Table 2 for description of the available Connectivity settings.
Touch
in General settings to enter Connectivity settings
Select appropriate communication protocol
ASTM HL
ASTM LL
HL7
POCT1-A
Communication protocol is disabled as default.
Select which results to be transferred to LIS/HIS/EMR by selecting the appropriate button
Patient results only
Patient and control results
Select "New results only" and previous obtained results will not be transferred to the LIS/HIS/EMR
Touch
to enter page 2 of the configuration or
Touch
to enter Server IP and Port number, Receiving application (available for ASTM HL, ASTM LL and HL7 only) and Receiving facility (available
for HL7 only).
Enter the server IP address: Press
Enter the server port number setting: Touch
Enter the Receiving application setting: Press
Use the button to select Patient ID as (available for HL7 only):
HIS Patient ID
Visit Number
Touch
to return to the General Settings menu
Table 2 Connectivity settings
Consult your network administrator and LIS/HIS/EMR administrator for required connectivity settings.
Protocol
ASTM HL
ASTM LL
HL7
POCT1-A
Disabled
Results
Patient results only
Patient and quality
control
Server IP
Port
Receiving
Application
Receiving Facility
Patient ID As
HIS Patient ID
Visit Number
For further information about the connectivity settings, see the Afinion 2 data sheets for POCT1-A, ASTM and HL7 which can be obtained at www.Alere.com or
through your local Afinion supplier.
18 | US
to return to the General settings menu
to continue to the Port number setting.
to continue to the Receiving application setting or to close the text input.
to continue to the Receiving facility setting or to close the text input.
ASTM High Level: The communication protocol is based on ASTM E 1394 - 97
ASTM Low Level: The communication protocol is based on ASTM E 1381 - 95
HL7: The communication protocol is based on HL7 version 2.4
POCT1-A: The communication protocol is based on CLSI: POCT01-A2 Point-of-Care Connectivity; Approved
Standard – Second Edition
Data connectivity is disabled
Only patient results will be transferred to the LIS/HIS/EMR
Both patient and QC results will be transferred to the LIS/HIS/EMR
Insert the IP address of the receiving system [0-255.0-255.0-255.1-254]
[0-65535] (0 = not set)
(ASTM HL, ASTM LL and HL7 only)
Receiving application name (0 – 30 characters)
(HL7 only)
Receiving facility name (0-30 characters)
(HL7 only)
(HL7 only)
AFINION
2 User Manual