Owner’s Manual Fluid Warming Cabinet DC250L DC400L 120V DC250L shown with single basket insert DC400L Enthermics Medical Systems An ISO 13485:2016 certified company W164 N9221 Water St | Menomonee Falls WI 53051 Tel 262-251-8356 | 800-TO-B-WARM generalinfo@enthermics .com www .enthermics .com Printed in the U.S.A.
. fi led and actively pursued . Enthermics cannot, however, 2 . Do not sign a delivery receipt or a freight bill until fi le any damage claims, assume the responsibility of any a proper count has been made and inspection of all claims, or accept deductions in payment for such claims .
. 4 . Slowly roll the warmer across high thresholds . Dimensions (HxWxD) DC250L: 21 .9" x 18 .4" x 25 .8" (557mm x 470mm x 630mm) DC400L: 21 .9" x 24 .0" x 27 .1" (557mm x 610mm x 688mm)
Safety Procedures • The fl uid warming cabinet is only intended for Knowledge of proper procedures is essential to the warming medical solutions for irrigation and safe operation of electrically and/or gas energized injection prior to use . Refer to the labeling of the equipment .
Installation WARNING DANGER To prevent serious personal injury, death, or To prevent serious personal injury, death, property damage: or property damage: Do not steam clean, hose down or flood the Do not use this warmer in the presence of interior or exterior with water or liquid solution flammable anesthetic mixtures (with air of any kind.
. This tag is permanently attached to the appliance . Power Requirements NEMA 5-15P DC250L: 120 V .A .C . — 60 Hz, 1 ph 15A - 125V Plug 0 .7 kW, 5 .8 Amps...
Electrical Information Guidance and Manufacturer’s Declaration The warmer requires special precautions may result if the power cord attached is altered or a regarding EMC (Electromagnetic manufacturer supplied power cable is not used . Compatibility) and needs to be installed The warmer should not be used adjacent to or stacked with and put into service according to the other equipment .
Page 9
Electrical Information Electromagnetic Emissions The warmer is intended for use in the electromagnetic environment specifi ed below . The customer or the end user of this warmer should assure that it is used in such an environment . Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the warmer, including...
Operation Instructions Cleaning Agents Always use the proper cleaning agent at the manufacturer’s recommended strength . Contact a local cleaning supplier for product recommendations . Preparation Before Operating the Warmer 1 . Remove and wash any detachable items such as the insert assembly with hot, soapy water .
Operation Instructions Using the Fluid Controller Temperature Range: • Irrigation (IRR) temperature range: 32°C to 66°C (90°F to 150°F) • Injection (INJ) temperature range: 32°C to 40°C (90°F to 104°F) How to Set the Temperature Scale How to Set the Operational Sound Volume 1 .
Page 12
Operation Instructions Using the Fluid Controller (continued) Temperature Range: • Irrigation (IRR) temperature range: 32°C to 66°C (90°F to 150°F) • Injection (INJ) temperature range: 32°C to 40°C (90°F to 104°F) How to Lock the Controller The controller can be locked to prevent changes being made to the temperature set-point or the mode selection .
8 . Make sure the warmer door is securely closed aft er initial loading and following each fl uid removal . Important Do not load the fluid warming cabinet beyond the recommended maximum capacity: DC250L DC400L 8 (1) liter bags per 14 (1) liter bags per...
Cleaning and Preventative Maintenance WARNING WARNING To prevent serious personal injury, death, or Electric shock hazard. property damage: Perform lockout/tagout procedures before Do not steam clean, hose down or flood the cleaning or servicing this appliance. interior or exterior with water or liquid solution of any kind.
Cleaning and Preventative Maintenance Preventative Maintenance Checklist Daily Checklist: Is the operation and care manual available? Has everyone been trained in the operation and safety instructions of this appliance? Do not overload the appliance . Make sure the air vents in the airfl ow insert panels located inside the fl uid chamber are not obstructed .
Troubleshooting NOTE: All non-critical codes can be cleared using the ON/OFF This section is provided for the assistance button. Critical errors (marked with a *) can only be of qualified and trained service cleared by pressing the circuit breaker at the rear of the technicians only and is not intended warmer to the off...
Page 19
• Call service . P131 Pad sensor 1 Heater pad over-temp error P231 Pad sensor 2 • Soft ware disengages heating pads . P331 Pad sensor 3 • Acknowledge error by pressing the ON/OFF button . Troubleshooting P431 Pad sensor 4 •...
Service How to Replace the Fuse WARNING For protection against fire and electrical shock use only UL listed 10A, 250V fast acting fuses, 5mm x 20mm (F1, F2). Access should be made by qualified service technicians only. FU-34955 FU-34955 F1 fuse A hospital grade cord must be used.
Service Full Assembly (DC250L shown) Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-39216 • Rev 3 • 12/18 • DC Series Fluid Warmer...
Page 23
Service Full Assembly (DC250L shown) Item Description DC250L DC400L Panel, door CV-35300 CV-35301 Door assembly, includes panel standard 5015297 5015298 Door assembly, with manual lock optional 5016415 5015875 Cam, tab lock LK-35651 LK-35651 Screw, cam, 1/4-20 x 1/2" FH SC-27556...
Page 24
Service Fan Assembly (DC250L shown) Item Description DC250L Qty DC400L FA-39130 FA-39130 Fan Wire, Health Monitor DC WI-39176 WI-39176 Fan Wire, Power Cord DC WI-39177 WI-39177 Thermostat, manual reset TT-35688 TT-35688 Spacer, 7/16", Nylon SP-39901 SP-39901 Spacer Support, 1/4", Nylon...
Page 25
Service Interface Assembly (DC400L shown) Item Description DC250L DC400L Screws, M4 x 0 .7 x 12mm plastic thread cutter SC-29333 SC-29333 Washer, Flat WS-22323 SC-29333 Washer, rubber, flat WS-25056 WS-25056 Panel, front top trim 18 .5" (470mm) 1011032 — —...
Page 26
Service Interior (DC250L shown) Item Description DC250L DC400L Door switch SW-35416 SW-35416 Top airflow insert IS-29880 IS-29881 Cavity sensor PR-37140 PR-37140 Sensor block BK-29882 BK-29882 Screw, M4 x 0 .7 x 6mm SC-22271 SC-22271 LED assembly (includes cover, ring and LED)
Wire Diagram Serial number is required for all inquiries. Part numbers and drawings are subject to change without notice. MN-39216 • Rev 3 • 12/18 • DC Series Fluid Warmer...
Page 29
Enthermics Medical Systems An ISO 13485:2016 certified company W164 N9221 Water St | Menomonee Falls WI 53051 Tel 262-251-8356 | 800-TO-B-WARM generalinfo@enthermics .com www .enthermics .com Printed in the U.S.A. Specifications are subject to change without notice. Made in the U.S.A.