Owner’s Manual ivNow ® 120V/220V/230V Enthermics Medical Systems An ISO 13485:2003 certified company W164 N9221 Water St | Menomonee Falls WI 53051 Tel 262-251-8356 | 800-TO-B-WARM generalinfo@enthermics.com www.enthermics.com Printed in the U.S.A. Specifications are subject to change without notice. Made in the U.S.A.
Serial number: ______________________________________________ information packed with the appliance . Purchased from: ______________________________________________ Indicates that the package contents should Date installed: ______________________________________________ not be used if the package has been damaged or opened. MN-38645 • Rev 1 • 04/18 • ivNow Fluid Warmer...
. Additional manuals are available from the Enthermics service department . 3 . Remove all protective plastic fi lm, packaging materials, and accessories from the appliance before connecting electrical power . MN-38645 • Rev 1 • 04/18 • ivNow Fluid Warmer...
NOTICE: For appliances delivered for use in any location regulated by the following directive ( 2012/19/EU -WEEE) Do not dispose of electrical or electronic appliances with other municipal waste. MN-38645 • Rev 1 • 04/18 • ivNow Fluid Warmer...
Installation How to Install/Mount the ivNow ivNow modular units can be installed in a variety of ways . Up to 3 units can be linked together with a jumper cord and connected via a mounting plate . Configurations can be placed on counter top, mounted to wall or pole .
Installation Mounting Instructions • Mount the rear plate to the wall using 1/4" toggle bolts and washers. Note: For drywall mounFng, drywall must be a minimum thickness of 1/2" • Remove the two (2) thumb screws from the sides of the rear bracket and set aside. • Li: the top bracket off of the assembly and set aside. • Remove the rear bracket from the assembly. • Use the panel as a template for the hole locaFons Ensure toggle anchors are fully Fghtened MN-38645 • Rev 1 • 04/18 • ivNow Fluid Warmer...
The power specifi cations are located on the unit identifi cation nameplate . This nameplate is permanently attached to the unit and must be located to verify power requirements . MN-38645 • Rev 1 • 04/18 • ivNow Fluid Warmer...
General Information Safety Feature The ivNow fluid warmer quickly warms and maintains • The control of the ivNow is designed to display an error the temperature of injection/intravenous solutions prior message (E-31) and stop heating if the temperature of to their use . The specially contoured warming module the fluid bag (as monitored by the dual sensor switch) cradles solution bags in 0 .5-, 1-, 2- &...
Page 10
5 . The display will continue to show the current bag temperature as the fluid is heated . 6 . The green status indicator will illuminate when the fluid is within +0/-2°C (+0/-3°F) of the setpoint temperature . MN-38645 • Rev 1 • 04/18 • ivNow Fluid Warmer...
Page 11
Fluid bag heated residence time is in excess of limit Temperature set point The DISPLAY will flash the word “dAtE” . ivNow will Press and hold the POWER button for 40 seconds or monitor the time spent continuously heating a fluid bag until the current set point (default of 40 °C or 104°F) is...
7 . Check that all display LEDs light up . Test by turning the unit off and then on . All LEDs illuminate for one second at start-up . 8 . Contact the Enthermics Service department for immediate repair if any problems exist . WARNING...
. 3 . If the warmer still displays a temperature, contact service . MN-38645 • Rev 1 • 04/18 • ivNow Fluid Warmer...
Page 14
Enthermics Medical Systems An ISO 13485:2003 certified company W164 N9221 Water St | Menomonee Falls WI 53051 Tel 262-251-8356 | 800-TO-B-WARM generalinfo@enthermics.com www.enthermics.com Specifications are subject to change without notice. Made in the U.S.A. Printed in the U.S.A.