9.2 Patient Error Messages

If an error message is displayed when the patient powers on the WatchPAT™ device, the
patient should take the actions specified below. If the problem persists the patient may
contact Itamar or an authorized representative directly.
Error
WatchPAT™ device
doesn't switch on
Probe disconnected
Hardware code
WatchPAT™300 System
Table 2 – Patient Troubleshooting
Possible Reason
ON button not activated
uPAT probe not connected
Probe may not be connected,
or may be a used probe
WatchPAT™ device failure
Action
Press the ON button firmly for at least 3
seconds
Ensure probe is connected and try again
Check connection of probe to cable and cable to
the WatchPAT™ device; check if probe has been
previously used and replace with new probe if
necessary
Contact Itamar or authorized representative
45
Operation Manual