Patient Error Messages; Table 2 - Patient Troubleshooting - Itamar Medical WatchPAT 200 Operation Manual

For use with patients suspected to have sleep related breathing disorders
Hide thumbs Also See for WatchPAT 200:
Table of Contents

Advertisement

9.2 Patient Error Messages

If an error message is displayed when the patient powers on the WatchPAT™ device,
the patient should take the actions specified below. If the problem persists the patient
may contact Itamar or an authorized representative directly.
Error
Oximetry sensor light
turns off while
WatchPAT™ device is on.
Oximetry sensor
disconnected
WatchPAT™ device does
not switch on
Probe disconnected
Hardware code
WatchPAT™200 System
Table 2 – Patient Troubleshooting
Possible Reason
Oximetry sensor plug not fully
inserted
Faulty oximetry sensor
Sensor not properly connected
or faulty
ON button not activated
®
PAT
probe not connected
Probe may not be connected,
or may be a used probe
WatchPAT™ device failure
Action
Verify that oximetry sensor plug is fully
inserted into the WatchPAT™ device
Check the oximetry sensor probe for damage
and replace if necessary
Check connection. If problem persists
replace sensor
Press the ON button firmly for at least 3
seconds
Ensure probe is connected and try again
Check connection of probe to cable and cable
to the WatchPAT™ device; check if probe has
been previously used and replace with new
probe if necessary
Contact Itamar or authorized representative