Warnings
*
This unit must be used with the guidance of a Physiotherapist or Doctor.
*
Type BF equipment, Continuous Operation.
*
Do not insert lead wires into a mains power supply.
*
Do not immerse unit into water or any other substance.
*
The unit is not protected from the ingress of shower of rain if used outside
the carrying case.
*
Do not use the NeuroTrac® Continence unit in the presence of a
flammable anaesthetic gas mixture.
*
Please use 9 Volt alkaline battery (IEC code: 6LR61, size: PP3). If using
rechargeable 9 Volt battery, be sure to use Nickel Metal Hydride type
batteries (IEC Code 6HR61).
We advise not to use Ni-Cad rechargeable batteries.
Caution: Do not use lithium batteries unless they comply with
IEC60086-4.
*
Patient Electrodes are for single patient use only.
*
Keep out of reach of children.
*
Do not use this stimulator on your facial area unless you are under strict
guidance from a qualified Clinician.
*
Application of electrodes near the thorax may increase the risk of cardiac
fibrillation.
*
Operation in close proximity to shortwave or microwave therapy
equipment may produce instability in the stimulator output.
*
Simultaneous connection of a patient to high frequency surgical equipment
may result in burns at the site of the stimulator electrodes and possible
damage to the stimulator.
*
This device can deliver current densities in excess of 2mA/cm2 when used
at a high intensity with small electrodes.
*
No modification of this equipment is allowed.
NeuroTrac
Continence Operation Manual
®
3