7.
Turn on the unit by pressing the on/off button once.
8.
Select the desired programme by pressing the PRG button (please refer to
the programme tables on page 12).
9.
Hold down the A+ button until you feel a strong but comfortable current.
The current may be decreased by pressing the A- button.
10. You may adjust the strength of the current during the treatment by pressing
the A+ or A- buttons.
11. To exit the programme before it is finished, turn off the unit by pressing the
on/off button once.
12. Important: Do not remove your probe / electrodes while the unit is running.
13. Carefully clean vaginal or rectal electrode before and after use. Wash
the probe gently in mild soapy water, rinse and make sure the probe is
completely dry before returning to storage in the plastic bag.
Low Battery Indicator
When the battery power is low, the low battery indicator will appear on the
screen (shown in the diagram on page 7). When the battery indicator shows one
bar, replace the battery.
Disconnection Indicator
When the probe becomes disconnected or when the lead wires do not conduct
the electrical current, the milliamp level will return to zero and the effected
channel will flash on and off.
Common Issue:
If you press + button but the intensity stops at 6-9 mA and cuts off to zero
and you see 0 mA flashing on the screen:
- This is a typical situation for any muscle stimulator. You are experiencing
the safety cut off due to no electrical load on the output. You should be able
to resolve this issue yourself, please follow the troubleshooting chapter on
page 21.
NeuroTrac
Continence Operation Manual
®
9