Multi-therapy pump for homecare patients (11 pages)
Summary of Contents for Hospira Plum 360
Page 1
Status Date Z09/EPS-97316-001/000/01 Approved Pend RL 01/08/2015 DCR Number 500012137 Document Description ARTWORK, MANUAL, TECHNICAL SERVICE, PLUM 360 Document Characteristics and Current Values Document Owner Owner Code Owning Site Functional Area Suzanne Willey SJRD00-SAN JOSE R AND D SJ-SAN JOSE USA RD-RESEARCH AND DEV.
Page 2
Compatible with Software For use with the following list numbers: 30010 30011 30012 (Including English for Canada list numbers 30010-65, 30011-65, and 30012-65) Technical Service Manual Hospira, Inc. 275 North Field Drive Lake Forest, IL 60045 USA 430-97316-001 (E, 2014-10)
Page 3
This page intentionally left blank. Plum 360 Infuser Technical Service Manual...
Page 4
(D, 2014-06) 430-97316-001 Changed Sections 1, 4, 5, 6, 7, 8, 9 to align content with current (E, 2014-10) infuser hardware/software. Removed Index. Updated front and back covers with current revision information and symbols. Technical Service Manual Plum 360 Infuser...
Page 5
CHANGE HISTORY This page intentionally left blank. Plum 360 Infuser Technical Service Manual...
If a problem in device operation cannot be resolved using the information in this manual, contact Hospira (see Section 6.1). for technical assistance Note: Servicing on the Plum 360 infuser is to be performed only by trained, authorized personnel. SCOPE This manual is organized into the following sections: Section 1 Introduction...
A MANDATORY ACTION SYMBOL MEANS THE INSTRUCTIONS THAT FOLLOW DESCRIBE A REQUIRED ACTION. FAILURE TO OBSERVE A MANDATORY ACTION COULD IMPACT USER OR PATIENT SAFETY. Note: A note highlights information that helps explain a concept or procedure. Plum 360 Infuser 1 - 2 Technical Service Manual...
DPM Digital pressure meter ECG Electrocardiogram EEG Electroencephalogram EEPROM Electrically erasable programmable read-only memory EMC Electromagnetic compatibility EMG Electromyogram EMI Electromagnetic interference ESD Electrostatic discharge ETO Ethylene oxide FPGA Field programmable gate array Technical Service Manual 1 - 3 Plum 360 Infuser...
Page 19
PWM Pulse width modulator RAM Random access memory rms Root-mean-square RTC Real-time clock SCC Serial communication controller SCP Serial communication port SLA Sealed lead acid SMT Surface mount technology SPI Serial peripheral interface Plum 360 Infuser 1 - 4 Technical Service Manual...
Microsecond USER QUALIFICATION The Plum 360 infuser must be used at the direction of or under the supervision of licensed physicians or certified healthcare professionals who are trained in the use of the infuser and the administration of parenteral and enteral fluids and drugs, and whole blood or red blood cell components.
SECTION 1 INTRODUCTION ELECTROMAGNETIC COMPATIBILITY The Plum 360 infuser (List Number 30010 with Module 30011 or 30012) has been tested to the requirements of IEC 60601-1-2:2007 and IEC 60601-2-24:2012. The infuser meets the EMC requirements of the Medical Device Directive 93/42/EEC with amendments by 2007/47/EC.
The user of the Plum infuser shall not perform any modifications to the wireless module other than the repairs described in Section 7.2.6. Unauthorized changes or modifications to the wireless system voids the user’s authority to operate the wireless system of the Plum infuser. Technical Service Manual 1 - 7 Plum 360 Infuser...
EQUIPMENT, THE INFUSER SHOULD BE OBSERVED TO VERIFY NORMAL OPERATION IN THE CONFIGURATION IN WHICH IT WILL BE USED. Note: The Plum 360 infuser does not require any special installation. The only preparation needed is for the biomedical technician to set the facility desired default settings and fully charge the battery.
MECHANISM INITIALIZATION IN PROGRESS prior to the CASSETTE TEST IN PROGRESS message. 6. A message may appear. Press NEW PATIENT? [YES] 7. Press to turn off the infuser. [ON/OFF] Technical Service Manual 1 - 9 Plum 360 Infuser...
KEYPAD STOP CONNECTED p q r s t u v w x y z INDICATOR AC ON ON OFF CLEAR LOCK KEYPAD Figure 1-1. Plum 360 Enabled Display and Keypad Plum 360 Infuser 1 - 10 Technical Service Manual...
Note: If the wireless icon or Ethernet icon does not appear in the lower right hand corner of the display, or the green Activity Indicator LED (see Figure 1-7) on the CE module is not illuminated periodically, contact the local IT representative or Hospira. Technical Service Manual 1 - 11 Plum 360 Infuser...
SECTION 1 INTRODUCTION 3. Check that the Hospira MedNet connectivity symbol is displayed above the wireless or Ethernet symbol. See Figure 1-5. Figure 1-5. Hospira MedNet is Active Symbol PUMPING Heparin 25,000 Units 250mL Dose units/kg/hr Rate mL/hr 499. Vol Inf...
1.12 BIOMED 1.12 BIOMED The BIOMED mode is used by qualified, trained biomedical engineering personnel to view and/or change settings in the Plum 360 infuser. The BioMed Settings page (see Figure 1-8) contains the following menu options: - Preventive Maintenance - Section 1.12.3...
INDICATOR AC ON ON OFF CLEAR LOCK KEYPAD ON/OFF Decimal button Figure 1-10. Plum 360 ON/OFF Button and Decimal Key The infuser performs a system self test. Plum 360 Infuser 1 - 14 Technical Service Manual...
Figure 1-11. System Self Test Screen 3. When the self test is complete, the Biomed Passcode screen appears. Passcode Enter Passcode for Biomed Passcode Use keypad then press Enter Enter Cancel Figure 1-12. Biomed Passcode Screen Technical Service Manual 1 - 15 Plum 360 Infuser...
Software Choose Request Figure 1-14. Biomed Settings Menu Note: The Software Request softkey only appears if the infuser is connected to the Hospira MedNet system. For instructions on updating the software, see Section 1.12.2. Plum 360 Infuser 1 - 16...
1.12 BIOMED 1.12.2 SOFTWARE UPDATES When new versions of the Plum software are available, Hospira notifies the Biomed technician. The Biomed technician then updates the software in the Plum infuser via the Hospira MedNet connection. To update the Plum software with the latest version: (see Section 1.12.1).
Turn on the infuser and access the Biomed mode. b. Select Display Logs on the Biomed Settings screen. c. Open Device Logs (see Section 1.12.6). d. Check the device log for the entry SWInstallSuccessful (Figure 1-19). Plum 360 Infuser 1 - 18 Technical Service Manual...
Additionally, the updated version number displays on the opening screen. Note: The infuser should not be placed into service if the software update has not been verified. 7. After re-booting the infuser, make sure the Hospira MedNet and Ethernet (or wireless) icons are displayed (see Section 1.11).
PREVENTIVE MAINTENANCE To set Preventive Maintenance notifications: 1. Use the navigation keys on the keypad to scroll to Preventive Maintenance in Biomed Settings and press Choose. Figure 1-20. Plum 360 Navigation Keys Biomed Settings Serial Number: 17705238 Preventive Maintenance Set Time and Date...
Figure 1-24. Clear Battery Warning Message 4. Press Yes to change the battery, and press No to retain the battery logs and return to the main menu. 5. Press Done to return to the main menu. Technical Service Manual 1 - 21 Plum 360 Infuser...
Note: Time zone offset is from Coordinated Universal Time UTC/GMT. Note: When the Time Zone offset is set with GMT, the time stamp in the Display Logs menu option displays the correct time of day for local time. Plum 360 Infuser 1 - 22 Technical Service Manual...
Note: If the device is connected to Hospira MedNet, the time is automatically set and the Biomed only has to set the time zone field. If the device is not connected to Hospira MedNet, the Biomed only has to set the local time; the Time Zone field will be of no use.
Pending until the system reboots. After exiting the Biomed mode and powering off, the software changes to the selected language upon the next reboot for either Biomed or clinical mode. Plum 360 Infuser 1 - 24 Technical Service Manual...
AC power are removed for a short or extended amount of time. The log data is stored in non-volatile memory. For a complete list of alarm codes and error codes, see Section 6.2.1 Section 6.2.2. Technical Service Manual 1 - 25 Plum 360 Infuser...
1. Use the navigation keys to scroll to Infuser Settings and press Choose: Biomed Settings Serial Number: 10162047 Preventive Maintenance Set Time and Date Manage Languages Display Logs Infuser Settings Manage Network Press Choose to Select Choose Figure 1-34. Infuser Settings Menu Option Plum 360 Infuser 1 - 26 Technical Service Manual...
Default End of Infusion: Choose KVO / Rate to change Press Done to save and return KVO/Rate Done Figure 1-36. Infuser Defaults Screen 3. Select the option and change the setting. Technical Service Manual 1 - 27 Plum 360 Infuser...
0.1 - 99.9 mL/hr and 100 - 999 mL/hr Allow standby Yes or No Maximum Standby Time 72 hours 24 - 72 hours in 1 hour increments Allow Delayed Start Yes or No Plum 360 Infuser 1 - 28 Technical Service Manual...
Previous Choose Screen Figure 1-39. Network Settings CE 3.0 Screen Note: The data displayed in the wireless, Ethernet, and Hospira MedNet Settings screens is sent from the CE (Connectivity Engine) as part of the communication protocol. Technical Service Manual 1 - 29...
Yes: Will Reset No: Cancel Figure 1-47. Reset Network Settings Screen 3. Press Yes to reset the network settings to the defaults or No to cancel and return to the previous screen. Technical Service Manual 1 - 33 Plum 360 Infuser...
CE and Plum! Yes: Shuts down CE then the Plum No: Cancel Figure 1-49. Confirm CE Shutdown Warning Screen 3. Press Yes to confirm the shutdown of the entire device. Plum 360 Infuser 1 - 34 Technical Service Manual...
Note: Wait at least five minutes for the CE module to power down and the microprocessor to save data before you remove the CE module. 4. Remove the CE module, if planning to replace it. For the replacement procedure, Section 7.2.6. Technical Service Manual 1 - 35 Plum 360 Infuser...
Page 51
SECTION 1 INTRODUCTION This page intentionally left blank. Plum 360 Infuser 1 - 36 Technical Service Manual...
(whether such cause be based in contract, negligence, strict liability, other tort, or otherwise) exceed the price of such product, and in no event shall Hospira be liable for incidental, consequential, or special damages or losses or for lost business, revenues, or profits.
Page 53
SECTION 2 WARRANTY This page intentionally left blank. Plum 360 Infuser 2 - 2 Technical Service Manual...
Section 3 SYSTEM OPERATING MANUAL A copy of the System Operating Manual is included with every Plum 360 infuser. If a copy contact Hospira (see Section 6.1). is not available, Technical Service Manual 3 - 1 Plum 360 Infuser...
Page 55
SECTION 3 SYSTEM OPERATING MANUAL This page intentionally left blank. Plum 360 Infuser 3 - 2 Technical Service Manual...
Section 4 THEORY OF OPERATION This section describes the theory of operation for the Plum 360 Infuser. The theory of operation details the general description, electronic subsystem overview, printed wiring assemblies, LCD, and mechanical overview of the infuser. GENERAL DESCRIPTION...
(SCC) channels and the serial communication port (SCP) channels. The 24 bit address bus is capable of accessing up to 16 MB of data. Plum 360 Infuser 4 - 2 Technical Service Manual...
DC/AC inverter. Intensity control is achieved by superimposing a DC control signal with the feedback signal. The DC control signal is sourced by a voltage divider consisting of a digitally controlled non-volatile potentiometer and three series diodes. Technical Service Manual 4 - 3 Plum 360 Infuser...
WDI remains either high or low for longer than the watchdog timeout period. After a reset, the software reads this memory-mapped bit to determine if the latest reset was a watchdog timeout. Plum 360 Infuser 4 - 4 Technical Service Manual...
The analog inputs are selected by the channel multiplexer according to the input address (see Table 4-1). The input multiplexer is a break-before-make type to reduce input-to-input noise injection resulting from channel switching. Technical Service Manual 4 - 5 Plum 360 Infuser...
BUZTST Buzzer test voltage LCDTST LCD contrast test voltage TUBTST CCFT intensity test voltage MI_STA Motor current A control MI_STB Motor current B control ) / 2 ref(+) ref(-) ref(-) ref(+) Plum 360 Infuser 4 - 6 Technical Service Manual...
(LEDAE*, LEDBE*). The AC power on LED indicates the status of AC power (LEDAC). A power supply AC on signal (BACON) controls the LED and is active only when AC power is connected. Technical Service Manual 4 - 7 Plum 360 Infuser...
Raw, unregulated charger voltage or battery voltage V2_7 2.7 V backup power for RTC and non-volatile SRAM Full time 5 V supply, backed up by supercap V12_0 12 V, low current supply for audio alarm Plum 360 Infuser 4 - 8 Technical Service Manual...
Flag, I/O valve cam sensor D, I FLLS_A Flag, L/S valve A pin detector D, I FLLS_B Flag, L/S valve B pin detector D, I FLPLRO Flag, plunger rotation sensor D, I Technical Service Manual 4 - 9 Plum 360 Infuser...
Page 65
Motor power is directly from power supply PWA V2_5 Reference voltage for ADC and DAC A, I Legend: P = Power A = Analog D = Digital I = Input O = Output Plum 360 Infuser 4 - 10 Technical Service Manual...
Resistor R70 and capacitor C46 filter the voltage across R71 and feed it back to the current sense input (1.5 V threshold) of U12. The duty cycle of U12 is reduced until the excessive load is removed. Technical Service Manual 4 - 11 Plum 360 Infuser...
STRTUP signal, via interrupt, to signal a user request for turning off the infuser. Figure 4-2 illustrates the system startup/shutdown sequence while battery powered. System power is always on while AC powered. Plum 360 Infuser 4 - 12 Technical Service Manual...
The software battery management algorithm controls the battery charger. The charging scheme is a current limit/two stage voltage limit charger. The charge current is limited to 1.3 A and the voltage is limited to either 6.74 V or 7.35 V Technical Service Manual 4 - 13 Plum 360 Infuser...
4.2.2.9 BATTERY The battery employed by the Plum 360 infuser is a sealed lead acid (SLA) type rated at 6 volts DC and a current capacity of at least 4.0 ampere-hours. A fully-charged, new battery provides three hours of operation at 125 mL/hr, or delivers 250 mL total volume, if pumping at a rate of 126 mL/hr or greater, whichever occurs first.
(phases) are completed. The step-angle (the number of steps per shaft revolution) resolutions are 3.6 degrees/step (100 steps/rev) for the plunger motor, and 7.5 degrees/step (48 steps/rev) for the I/O and L/S valve motors. Technical Service Manual 4 - 15 Plum 360 Infuser...
Optical switch flag sensors are used for tracking: - Plunger motor rotational position (coupler flag) - Plunger translational (linear) position - I/O valve motor rotational position (cam flag) - L/S valve motor rotational position (cam flag) Plum 360 Infuser 4 - 16 Technical Service Manual...
(proximal) or out of (distal) the cassette. Both sensors are piezoelectric crystal transmitter receiver pairs. Liquid between the transmitter and receiver will conduct the ultrasonic signal, while air will not (see Figure 4-4). Technical Service Manual 4 - 17 Plum 360 Infuser...
V2_5 supply. DIARE and PXARE are used to enable the distal and proximal sensors, respectively. The detector stage consists of an emitter follower, charging a 400 microsecond time constant, refreshed every 40 microseconds (twice per VCO sweep). Plum 360 Infuser 4 - 18 Technical Service Manual...
The FSR is arranged in a voltage divider configuration with a fixed resistor, followed by a comparator with hysteresis. The comparator circuits are located on the CPU PWA. The comparators are designed to trip as the FSR’s resistance falls below 120 KΩ. Plum 360 Infuser 4 - 20 Technical Service Manual...
4.2.4 CONNECTIVITY ENGINE MODULE REVISION 3.0 The Connectivity Engine Module Revision 3.0 contains: • Memory to be used by the Plum 360 infuser CPU • CE 3.0 Subsystem to connect to Hospira MedNet software • Nurse Call interface (see Figure 4-6) •...
Page 77
RXD1/ LED Drivers Yellow CTS1/ Spare RTS1 RJ-11 Connector Nurse Call Nurse Call NURSE Isolation Circuit LOTSW Audio Volume BUZADJ/ Control BUZL Figure 4-7. Single Peripheral Assembly with CE 3.0 Subsystem Plum 360 Infuser 4 - 22 Technical Service Manual...
It also contains transceivers for Ethernet. The peripheral assembly contains the connector that serves as the interface to the Plum 360 infuser CPU PWA. In addition, the peripheral assembly contains the interface for Ethernet, Wi-Fi (802.11 a/b/g/n), Nurse Call, and Volume Control. Communication with the Hospira MedNet system requires both assemblies to be present and mated.
CPU accesses via the 96-pin board to board connector interface. They are used to support the various functions on the Plum 360 infuser and are powered from a 3.3 V supply provided by the CPU PWA. The 3.3 V...
4.2.4.8 iMX534 APPLICATION PROCESSOR The main processor that enables communication with the Hospira MedNet system is the Freescale iMX534. It has an 800 MHz 32-bit ARM Cortex-A8 CPU with 32 KB instruction and data caches. To support the CE 3.0 functionality, the chip has interfaces for sending and receiving Wi-Fi and Ethernet data, and is powered by the Freescale MC34708 Power Management IC.
The switch PWA contains the plunger translation position sensor that is one of six position sensors in the system. The switch PWA is located at the side of the infusion mechanism sub-chassis, and connects to the driver PWA. Plum 360 Infuser 4 - 28 Technical Service Manual...
Similarly, a second air-in-line sensor located distal to the pumping chamber initiates an audible alarm if a predetermined amount of air is detected. Plum 360 Infuser 4 - 30 Technical Service Manual...
0.33 mL of fluid per cycle. The plunger motion is synchronized to the valve motors to provide controlled fluid delivery. Figure 4-11 for infusion mechanism valve pins and sensor locations. Technical Service Manual 4 - 31 Plum 360 Infuser...
Figure 4-11. infusion Mechanism Valve Pins and Sensor Locations 4.5.2.3 INLET/OUTLET VALVE SUBSYSTEM The inlet/outlet valve subsystem is similar in function and build to the A/B valve subsystem (see Section 4.5.2.2). Plum 360 Infuser 4 - 32 Technical Service Manual...
During operation, if the screw/coupler flag passes through the interrupter module at the incorrect time sequence, a motor phase loss is detected. Technical Service Manual 4 - 33 Plum 360 Infuser...
Page 89
SECTION 4 THEORY OF OPERATION This page intentionally left blank. Plum 360 Infuser 4 - 34 Technical Service Manual...
WARNING COULD RESULT IN ELECTRICAL SHOCK. CLEANING AND DISINFECTION The following sections describe how to clean and disinfect the Plum 360 infuser between each patient use. These instructions should also be used to prepare the infuser for storage, repair, or preventive maintenance.
CAUTION: Do not spray cleaning solutions toward any opening in the infuser. CAUTION: Use all cleaning solutions as specified by the manufacturer, to avoid infuser damage. Use only recommended cleaning solutions and follow manufacturer’s recommendations regarding dilution. Using cleaning solutions not recommended by Hospira may result in product damage. 5.2.1...
Note: Administration sets must be removed by authorized personnel and disposed of per facility policy before cleaning the infuser. Note: If the minipole is attached, see Section 7.2.12 for instructions on removing the minipole assembly. Technical Service Manual 5 - 3 Plum 360 Infuser...
8 / surface 7 / surface 1 / surface 5 / surface 15 / surface Ensure the surfaces remain moist the entire duration of each time interval (depending on the procedure). Plum 360 Infuser 5 - 4 Technical Service Manual...
3. Wipe bottom surface using cleaning solution and lint-free cloths or wipes. 4. Lift the infuser, wipe the surface where it was resting, and place it upright. 5. Wipe top and side surfaces using cleaning solution and lint-free cloths or wipes. Technical Service Manual 5 - 5 Plum 360 Infuser...
Page 95
5-2). Note: Do not allow fluid into any ports or vents on the CE Module. 5. Wipe areas around the power cord outlet, the power cord, and the plug, including prongs. Plum 360 Infuser 5 - 6 Technical Service Manual...
For (see Figure 5-4) this reason, the cassette receptacle should only be cleaned by personnel who are aware of how the infuser operates and can take proper precautions. Technical Service Manual 5 - 7 Plum 360 Infuser...
Page 97
To clean the cassette receptacle: 1. Open the cassette door, and press the yellow-marked, door release tab as shown Figure 5-4. 2. Gently press the cassette door down until it opens completely. Plum 360 Infuser 5 - 8 Technical Service Manual...
Page 98
Note: If the roller does not spin smoothly, replace the door (see Section 7.2.13.8). 6. Clean the door lever, and use swabs to clean the crevices around it. Note: Do not allow fluid into the internal part of the infuser. Technical Service Manual 5 - 9 Plum 360 Infuser...
3. When no moisture is detected, close the cassette door. Note: Allow to air dry completely before beginning disinfection. Note: After the cassette door receptacle is dry, close the cassette door to avoid damage when moving the infuser. Plum 360 Infuser 5 - 10 Technical Service Manual...
Do not exceed 1 part bleach in 10 Household Bleach parts water (8.25% concentration) Note: At the time of publication, Hospira recommends only the disinfecting solutions in Table 5-4. For updated listings of approved disinfectants, visit www.hospiraparts.com. 5.3.2 DISINFECTING SUPPLIES...
Table 5-6. Minimum Time Intervals for Disinfecting Disinfectant in Contact with Infuser Rinsing with Water 15 minutes 3 minutes Ensure the surfaces remain moist the entire duration of each time interval (depending on the procedure). Plum 360 Infuser 5 - 12 Technical Service Manual...
2. Carefully swab the pins, sensors, and surrounding areas without tilting the infuser (see Figure 5-5). back 3. Use a swab dipped in 70% isopropyl alcohol to disinfect the air-in-line detectors and remove residue from cleaning. Technical Service Manual 5 - 13 Plum 360 Infuser...
3. When no moisture is detected, reattach the power cord Velcro strap. 4. Return the infuser to service. Note: Make sure the power cord and plug are completely dry before plugging the infuser into AC (mains) power. Plum 360 Infuser 5 - 14 Technical Service Manual...
5.4 PREVENTIVE MAINTENANCE PREVENTIVE MAINTENANCE Hospira requires that preventive maintenance be performed at least once every 12 months. Replace components as required by visual inspection and test results. Complete the Preventive Maintenance Checklist in Section 5.4.13. Note: Perform the Performance Verification Test along with the visual inspections as part of the preventive maintenance process at least once every12 months.
- Front Label - - Tubing Guide Consult Labels (A/B Labels) Accompanying Documents - (lower under keypad) 3. Inspect the labels for legibility and peeling. To replace a label, contact Hospira. Plum 360 Infuser 5 - 16 Technical Service Manual...
Page 106
MAC ADDRESS LABELS (ETHERNET AND WIRELESS) BRAND LABEL (PAD-PRINTED) PRODUCT I.D. LABEL (DRIVER) FOLLOW SERVICE POWER CORD INSTRUCTIONS REVISION LEVEL LABEL FOR USE LABEL LABEL (DRIVER) Figure 5-6. Plum 360 Infuser Labels Technical Service Manual 5 - 17 Plum 360 Infuser...
JACK SCREW STEP 5 6-32 x 5/8 PAN HEAD SCREW AC POWER CORD STEP 4 RETAINER AC POWER CORD POWER CORD LABEL Figure 5-7. Plum 360 Power Cord, Retainer, and Velcro Strap Plum 360 Infuser 5 - 18 Technical Service Manual...
If the door does not open smoothly, check for debris or dried fluid buildup. Clean the infusion mechanism as described in Section 5.2. 2. Move the door lever to the CLOSED position. Confirm smooth operation as described in Step 1. Technical Service Manual 5 - 19 Plum 360 Infuser...
7.2.13.8. smoothly, replace the door as described in 4. Confirm that the retaining ring, which secures the roller wheel to the pin, is seated properly, and the pin is not bent. Plum 360 Infuser 5 - 20 Technical Service Manual...
Figure 5-9. Door Roller Inspection 5.4.5 FLUID SHIELD INSPECTION Visually inspect the Plum 360 infuser fluid shield at least once every 12 months. Equipment required for the fluid shield inspection is a 0.025 inch (0.65 mm) feeler gauge (plastic or metal),...
Page 111
If the diaphragm is torn, cracked, or missing, contact Hospira for repair. CLOSE LEVER WHEN Figure 5-10. Releasing the Cassette Door FEELER GAUGE Figure 5-11. Inspecting the Fluid Shield with Feeler Gauge Plum 360 Infuser 5 - 22 Technical Service Manual...
2. Release the door so that it lays flat. Press the door release tab to the right and open the cassette door all the way (see Figure 5-10). 3. Inspect the distal pressure pin to determine that it is not damaged or broken. Technical Service Manual 5 - 23 Plum 360 Infuser...
2. Release the door so that it lays flat. Press the door release tab to the right and open the cassette door all the way (see Figure 5-10). 3. Inspect the pin to determine that it is not damaged or broken. Plum 360 Infuser 5 - 24 Technical Service Manual...
5-15, and proceed as follows: 1. Inspect for missing or loose rubber foot pads, and rubber foot pads that are starting to peel away from the enclosure. Replace or install as described in Section 7.2.3. Technical Service Manual 5 - 25 Plum 360 Infuser...
SECTION 5 MAINTENANCE AND SERVICE TESTS RUBBER FOOT PAD (5) Figure 5-15. Rubber Foot Pads on Plum 360 Infuser 5.4.9 POLE CLAMP INSPECTION AND TEST Visually inspect and test the pole clamp at least once every 12 months. To inspect the pole clamp, see...
Page 116
5.4 PREVENTIVE MAINTENANCE CHECK FOR SECURE ATTACHMENT TO REAR ENCLOSURE Figure 5-16. Pole Clamp and Extrusion on Plum 360 Infuser RUBBER PAD THREADED INTERFACE SHAFT TIP Figure 5-17. Pole Clamp Assembly Technical Service Manual 5 - 27 Plum 360 Infuser...
Contact Hospira to obtain a replacement battery. If spare parts, other than Hospira approved spare parts, are used for battery replacement, the warranty on the Plum 360 infuser shall be void. Note: The battery compartment must be visually inspected once every 12 months.
Page 118
15. Press to turn off the infuser. [ON/OFF] BATTERY WITH WIRE HARNESS ASSEMBLY BATTERY DOOR PAD SOFT PUSH ON RETAINING RING BATTERY DOOR SLOTTED HEX HEAD SCREW Figure 5-18. Removing the Battery Technical Service Manual 5 - 29 Plum 360 Infuser...
STOP CONNECTED p q r s t u v w x y z INDICATOR AC ON ON OFF CLEAR LOCK KEYPAD Figure 5-19. Plum 360 Infuser Keypad Plum 360 Infuser 5 - 30 Technical Service Manual...
(see Section 7.2.13.5). 8. Observe the display area. Confirm that the display is clear and readable. (see Section 7.2.13.2). 9. If any of the LEDs are not operating, replace the keypad Technical Service Manual 5 - 31 Plum 360 Infuser...
Page 121
KEYPAD STOP CONNECTED p q r s t u v w x y z INDICATOR AC ON ON OFF CLEAR LOCK KEYPAD Figure 5-20. Plum 360 Infuser Display and Indicators Plum 360 Infuser 5 - 32 Technical Service Manual...
Section 5.5.12, Distal Air-In-Line Test Section 5.5.13, Distal Occlusion Test Section 5.5.14, Delivery Accuracy Test Section 5.5.15, Nurse Call Test Section 5.5.16, Electrical Safety Test Section 5.5.17, End of the PVT Plum 360 Infuser 5 - 34 Technical Service Manual...
Proximal Air-in-Line test, a Distal Air-in-Line Test Setup Distal that is used only for the Distal Air-in-Line test, and a Occlusion Test Setup, which is a modification of the Basic test setup. Technical Service Manual 5 - 35 Plum 360 Infuser...
STOP p q r s t u v w x y z AC ON ON OFF CLEAR LOCK KEYPAD Figure 5-21. Basic Test Setup Plum 360 Infuser 5 - 36 Technical Service Manual...
Page 126
PlumSet with liquid (prime it), eliminating all air, and then load the cassette into the infuser. 1. Place the infuser on a bench or attach it to an IV pole. Technical Service Manual 5 - 37 Plum 360 Infuser...
Page 127
Note: Do not position the container above the infuser while inserting the piercing pin. Figure 5-24. Inserting the Piercing Pin 5. Suspend the container on an IV pole. 6. Check for leaks. If any part of the container is leaking, replace it. Plum 360 Infuser 5 - 38 Technical Service Manual...
Page 128
10. Tap and clear air from the cassette, Y-site, and tubing to remove all air from the remainder of the administration set (see Figure 5-28). Figure 5-28. Removing Air from the Administration Set Technical Service Manual 5 - 39 Plum 360 Infuser...
Page 129
1. Lift the lever to open the cassette door (see Figure 5-30). Figure 5-30. Opening the Cassette Door 2. Grasp the cassette by the finger grip (see Figure 5-31). FINGER GRIP Figure 5-31. Cassette Finger Grip Plum 360 Infuser 5 - 40 Technical Service Manual...
Page 130
7. Ensure that the score mark on the drip chamber is 12 to 24 inches higher than the cassette. 8. Place the distal end of the tubing in the collection container. Go to Section 5.5.2.1.4, to attach and prime the secondary line. Technical Service Manual 5 - 41 Plum 360 Infuser...
Page 131
4. Squeeze the drip chamber to fill it about 1/3 full or to the score mark (see Figure 5-34). Do not completely fill the drip chamber. Figure 5-34. Squeezing the Drip Chamber Plum 360 Infuser 5 - 42 Technical Service Manual...
Page 132
Clave. Move the Option-Lok collar over the Clave and twist (see Figure 5-37). clockwise to secure the line to the port OPTION-LOK COLLAR CLAVE SECONDARY PORT Figure 5-37. Option-Lok Collar and Clave Secondary Port Technical Service Manual 5 - 43 Plum 360 Infuser...
Page 133
8. Arrange the fluid container so that the score mark on the drip chamber is 12 to 24 inches higher than the cassette. Note: The secondary container does not need to be higher than the primary container for accurate delivery of a piggyback infusion. Plum 360 Infuser 5 - 44 Technical Service Manual...
Page 134
STOP p q r s t u v w x y z AC ON ON OFF CLEAR LOCK KEYPAD Figure 5-39. Complete Basic Test Setup Technical Service Manual 5 - 45 Plum 360 Infuser...
Figure 5-40. Preparing the Proximal Run-In Cassette 2. Using the permanent marker, write “Proximal” and the date on the drip chamber. 3. Follow the instructions for priming a run-in cassette assembly in Section 5.5.2.4. Plum 360 Infuser 5 - 46 Technical Service Manual...
Figure 5-41. Preparing the Distal Run-In Cassette 2. Using the permanent marker, write “Distal” and the date on the drip chamber. Section 5.5.2.4. 3. Follow the instructions for priming a run-in cassette assembly in Technical Service Manual 5 - 47 Plum 360 Infuser...
4. Turn on the infuser. During the self test, the infuser may issue a cassette test failure alarm. 5. Press and hold [BACKPRIME] to pump water from the drip chamber into the proximal lines and cassette. Plum 360 Infuser 5 - 48 Technical Service Manual...
Page 138
If there is any leakage, replace the run-in cassette. 4. Keeping the cassette upright, remove the white cap. 5. Pull out the flow regulator (see Figure 5-43). Figure 5-43. Pulling Out the Flow Regulator Technical Service Manual 5 - 49 Plum 360 Infuser...
Page 139
8. Release the pumping chamber and flow regulator. 9. Repeat Steps 5 through 8 until all distal air is pumped out of the tubing. 10. Replace the cap. The run-in cassette is now ready for use. Plum 360 Infuser 5 - 50 Technical Service Manual...
• Required equipment as listed for the Basic test setup in Section 5.5.2.1. • Three-way stopcock • Digital Pressure Meter (DPM) • Cloth or paper towel to catch drips. (Small amounts of water may be released from the stopcock during testing.) Technical Service Manual 5 - 51 Plum 360 Infuser...
Page 141
2. Attach the pressure sensor connector on the DPM to a compatible port (male or (see Figure 5-48). female) on the three-way stopcock 3-WAY STOPCOCK PORT DPM CONNECTOR Figure 5-48. Attaching the Three-Way Stopcock to the DPM Plum 360 Infuser 5 - 52 Technical Service Manual...
Page 142
AC ON ON OFF CLEAR LOCK KEYPAD UNIVERSAL PRESSUREMETER 0.00 0 to 12” cmOFHO INCHESOFHO mmHg -13.5TO15 -13.5TO75 PRESSUREINPUT Figure 5-50. DPM Connector Height Technical Service Manual 5 - 53 Plum 360 Infuser...
[ENTER]. 4. After the cassette test is complete, verify that CASSETTE TEST FAILURE is flashing on the display and that the alarm sounds. 5. Open the door and remove the cassette. Plum 360 Infuser 5 - 54 Technical Service Manual...
1. While the infuser displays the delivery screen, press [A] to select Line A. 2. Verify that the PROGRAM screen is displayed. Enter a rate of 23 mL/hr. Press (C) to clear the value. Technical Service Manual 5 - 55 Plum 360 Infuser...
Page 145
. Highlight VOLUME INFUSED and press [CHOOSE]. 17. Press [SETTINGS/VOLS/CCA] 18. Press the [CLEAR A]. Line A volume changes to 0 mL and the display returns to the previous screen. Plum 360 Infuser 5 - 56 Technical Service Manual...
Cancel Figure 5-51. Passcode to Lock Keypad To unlock the keypad: 1. Press any key except [STOP]. 2. When the Passcode dialog box appears, enter the passcode .963, and press Enter. Technical Service Manual 5 - 57 Plum 360 Infuser...
5. Verify that fluid is pumping; the message PUMPING is displayed in the Line A status bar, and the LED above Line A flashes. 6. Verify that the alarm sounds and the alarm message displays. Plum 360 Infuser 5 - 58 Technical Service Manual...
5. Press [AUDIO PAUSED] and verify that the alarm stops while the message on the display continues to flash. 6. Unclamp the proximal line and press [START]. Verify that pumping resumes. 7. Press [STOP]. 8. Open the cassette door and remove the cassette. Technical Service Manual 5 - 59 Plum 360 Infuser...
6. Before 1 mL of fluid is delivered, verify that pumping stops, the alarm sounds and the alarm message is flashing on the display. 7. Open the cassette door and remove the test cassette. Plum 360 Infuser 5 - 60 Technical Service Manual...
12. Verify that the distal pressure limit is set at 6 psi. If the pressure limit is not 6 psi, highlight the Distal Pressure Limit and enter 6. 13. Press [DONE]. 14. Press [PREVIOUS SCREEN]. Technical Service Manual 5 - 61 Plum 360 Infuser...
Page 151
NEW PATIENT? prompt. 19. Press [SETTINGS/VOLS/CCA] to select the SETTINGS/VOLS/CCA screen. 20. Select PRESSURE and press [CHOOSE]. 21. Select Distal Pressure Limit. Enter 10 psi, and press [DONE]. 22. Press [PREVIOUS SCREEN]. Plum 360 Infuser 5 - 62 Technical Service Manual...
Page 152
28. Open and close the cassette door to clear the alarm. 29. Turn off the Infuser. 30. Remove the distal tubing from the three-way stopcock and turn off the DPM. Technical Service Manual 5 - 63 Plum 360 Infuser...
AC ON ON OFF CLEAR LOCK KEYPAD 25 mL GRADUATED CYLINDER Figure 5-59. Delivery Accuracy Test Setup 2. Turn on the infuser. Plum 360 Infuser 5 - 64 Technical Service Manual...
To perform the nurse call test, proceed as follows: 1. Attach the 2-prong lead on the nurse call test cable to the ports on the DMM that are marked for measuring resistance. Technical Service Manual 5 - 65 Plum 360 Infuser...
Page 155
DMM (approximately 1 Ω on a scale of 0 to 100 Ω). If the short circuit appears, the test is successful. 8. Press [STOP], and turn off the infuser. Plum 360 Infuser 5 - 66 Technical Service Manual...
[ENTER]. 3. In response to the NEW PATIENT? prompt, press [YES]. 4. Remove the cassette and close the cassette door. 5. Turn off the infuser. 6. Return the infuser to service. Plum 360 Infuser 5 - 68 Technical Service Manual...
For technical assistance, product return authorization, and to order parts, accessories, or manuals from outside the United States, contact the nearest Hospira sales office. ALARM MESSAGES AND ALARM CODES Under most alarm conditions the infuser ceases normal operation, generates an audible alarm, and displays an alarm message or alarm code on the LCD screen.
[CANCEL] softkey. and both lines are delivering by pressing the STOP hardkey, but has not selected a line (A, B, or A&B) or selected Cancel to complete the action. Plum 360 Infuser 6 - 2 Technical Service Manual...
Page 160
Press any hardkey except N101 Start or clear lines. minutes when both lines [AUDIO PAUSED]. have been stopped by the user and not cleared or restarted. Note: Will reassert if the condition persists. Technical Service Manual 6 - 3 Plum 360 Infuser...
Page 161
CCA! the line was programmed or open the cassette door. Press STOP key under a different CCA than the CCA currently used; and Line B has a concurrent delivery was programmed. Plum 360 Infuser 6 - 4 Technical Service Manual...
Page 162
N185 A Startup! Open/close on Line A during the cassette Either backprime or open door or Backprime integrity test: (peak detection and close the cassette algorithm door. Check the syringe size. Technical Service Manual 6 - 5 Plum 360 Infuser...
Page 163
Line B (peak detection for kinks and correct any algorithm) during delivery found. Restart Line B. Note: The alarm can also be cleared by clearing the confirmed program or opening the cassette door. Plum 360 Infuser 6 - 6 Technical Service Manual...
Page 164
Distal AIR! A single air bolus is detected Open the cassette door. N234 Disconnect/reprime. at the distal sensor that Clear programs. Press START. exceeds the air detection threshold. Technical Service Manual 6 - 7 Plum 360 Infuser...
Page 165
1. Remove infusion N255 mechanism 2. Remove fluid shield 3. Visually inspect regulator closer 4. Verify regulator closer is completely engaged 5. Test If unrestricted flow persists, replace the infusion mechanism Plum 360 Infuser 6 - 8 Technical Service Manual...
Page 166
Other hardware failure causing excessive current Replace CPU PWA. draw. (see Section 7.2.13.4) Replace infusion mechanism assembly. (see Section 7.2.13.6) Reset time and date, if necessary. (see Section 1.12.4) Technical Service Manual 6 - 9 Plum 360 Infuser...
AC power and placing it into Standby mode (see the Plum 360 Infuser Operating Manual). If the E321 error code does not reappear after the charging cycle, the...
Page 168
Reset time and date, if necessary. (see Section 1.12.4) Power Off then On. Display Controller IC error. Power off the infuser. E342 Replace pump if Replace CPU PWA. alarm continues. (see Section 7.2.13.4) Technical Service Manual 6 - 11 Plum 360 Infuser...
Page 169
(see Section 7.2.13.4) Power Off then On. Generic plunger motor error E380 Replace pump if Replace infusion mechanism alarm continues. assembly. (see Section 7.2.13.6) Reset time and date, if necessary. (see Section 1.12.4) Plum 360 Infuser 6 - 12 Technical Service Manual...
Page 170
Reset time and date (see Section 1.12.4) E439 Power Off then On. Any key is held or activated for Power off the infuser. Replace pump if at least 2 consecutive minutes. alarm continues. Technical Service Manual 6 - 13 Plum 360 Infuser...
Page 171
(see Section 7.2.13.4) Reset time and date, if necessary. (see Section 1.12.4) Power Off then On. Volume infused and program Power off the infuser E451 Replace pump if delivery rate mismatch alarm continues. Plum 360 Infuser 6 - 14 Technical Service Manual...
E431 E447 Note: Some alarm codes include sub-ID codes. These sub-ID codes are intended for Hospira internal use only, and should be included when contacting Hospira. TROUBLESHOOTING PROCEDURES This section details recommended procedures for problems not associated with alarms requiring technical service.
Replace infusion mechanism assembly (see Section 7.2.13.6) Defective special cassette Replace special cassette Proximal Air-in-Line Test (Section 5.5.11) Dirty sensors Clean sensors Defective APP PWA Replace infusion mechanism assembly (see Section 7.2.13.6) Plum 360 Infuser 6 - 16 Technical Service Manual...
Page 174
Damaged or faulty cassette Replace cassette Defective infusion mechanism Replace infusion mechanism assembly assembly (see Section 7.2.13.6) Defective AC power cord Replace AC power cord Electrical Safety Test (Section 5.5.16) (see Section 7.2.5) Technical Service Manual 6 - 17 Plum 360 Infuser...
4. If the regulator closer is completely seated, reassemble the device and perform the PVT in Section 5.5. Regulator closer not seated correctly Regulator closer seated correctly Figure 6-1. Regulator Closer Plum 360 Infuser 6 - 18 Technical Service Manual...
Note: The device will internally network reset. It may take up to five minutes. 9. Verify that the infuser network is now set to the factory defaults, and proceed (see Table 1-2). to a normal network configuration of the device Technical Service Manual 6 - 19 Plum 360 Infuser...
Note: The device will internally reset. It may take up to five minutes. Note: Verify that the infuser is now set to the factory defaults, and proceed to a normal configuration of the device. Plum 360 Infuser 6 - 20 Technical Service Manual...
BEFORE PERFORMING REPLACEMENT PROCEDURES. CAUTION: Use proper ESD grounding techniques when handling components. Wear an antistatic wrist strap and use an ESD-protected workstation. Store PWAs in antistatic bags before placing on any surface. Technical Service Manual 7 - 1 Plum 360 Infuser...
Replacement of a rubber foot pad is routine maintenance and no verification procedure is normally required. However, if there is a possibility the infuser may have been damaged during these procedures, perform the PVT in Section 5.5. Plum 360 Infuser 7 - 2 Technical Service Manual...
Section 5.4.10. Use only Hospira-approved replacement batteries. Note: Contact Hospira to obtain a replacement battery. If spare parts, other than Hospira approved spare parts, are used for battery replacement, the warranty on the Plum 360 Infuser shall be void. WARNING: DISCONNECT THE INFUSER FROM AC POWER PRIOR TO OPENING THE THE DEVICE AND CHANGING THE BATTERY.
Page 181
10. Replace the battery door using the screw that was removed in step 3. 11. Press [ON/OFF] with the infuser disconnected from AC power, and verify the front panel battery symbol illuminates. Plum 360 Infuser 7 - 4 Technical Service Manual...
Page 182
PVT in Section 5.5. BATTERY WITH WIRE HARNESS ASSEMBLY BATTERY DOOR PAD SOFT PUSH ON RETAINING RING BATTERY DOOR SLOTTED HEX HEAD SCREW Figure 7-2. Battery Assembly Technical Service Manual 7 - 5 Plum 360 Infuser...
Replacement of the AC power cord, retainer, and Velcro strap is routine maintenance and no verification procedure is normally required. However, if there is a possibility the infuser may have been damaged during the procedure, perform the PVT in Section 5.5. Plum 360 Infuser 7 - 6 Technical Service Manual...
CE instructions in Section 1.12.8.5. Note: After disconnecting from AC power, wait at least five minutes for the CE to power down and the microprocessor to save data, then proceed to step 2. Technical Service Manual 7 - 7 Plum 360 Infuser...
Page 185
10. Perform the connectivity check in 11. Perform the PVT in Section 5.5 to verify successful CE module replacement. CE MODULE 4-40 X 5/8 PAN HEAD SCREW WITH WASHER (2) Figure 7-4. CE Module Plum 360 Infuser 7 - 8 Technical Service Manual...
Page 186
LOWER EMI NUT (2) SPRING SHIELD 4-40 x 5/8 #4 FLAT TORX KNOB WASHER, SCREW (4) COVER CE MODULE BLACK COVER ROUND SEAL, PURPLE I/O COVER COVER Figure 7-5. CE Module Technical Service Manual 7 - 9 Plum 360 Infuser...
7.2.4. 3. Carefully set the infuser face down. 4. Remove the CE module as described in Section 7.2.6. 5. Remove the antenna cover with a small screwdriver starting on the bottom. Technical Service Manual 7 - 11 Plum 360 Infuser...
9. Remove the screws that secure the peripheral assembly to the CE module cover and EMI shield. 10. Note the position of the two hex nuts installed in the PWA mounting brackets located on the CE module cover. Retain the nuts for reassembly. Plum 360 Infuser 7 - 12 Technical Service Manual...
9. Using the Phillips screwdriver, remove the screws in the infuser handle. 10. Using the flat blade screwdriver, press the flex tabs (see Figure 7-1) while lifting up the front enclosure, and remove the enclosure. Technical Service Manual 7 - 13 Plum 360 Infuser...
Page 191
POWER CORD 4-40 X 7/16 RETAINER SCREW JACK EQUIPOTENTIAL TERMINAL 6-32 X 3 1/4 MAIN PHILLIPS CHASIS SCREW (2) FRONT ENCLOSURE Figure 7-8. Separating the Front Enclosure, Main Chassis, and Rear Enclosure Plum 360 Infuser 7 - 14 Technical Service Manual...
9. Reinstall the battery and connect the infuser to AC power. 10. Perform the PVT in Section 5.5 to verify successful front enclosure, rear enclosure, or main chassis replacement. Technical Service Manual 7 - 15 Plum 360 Infuser...
6. Join the front enclosure, rear enclosure, and main chassis in the exact reverse order of separation. 7. Reinstall the battery and connect the infuser to AC power. 8. Perform the PVT in Section 5.5 to verify successful front/rear enclosure gasket replacement. Plum 360 Infuser 7 - 16 Technical Service Manual...
Page 194
7.2 REPLACEMENT PROCEDURES FRONT ENCLOSURE 4-40 X 1/4 PHILLIPS SCREW (2) PROXIMAL TUBING HOLDER FLUID SHIELD GASKET GASKET DISPLAY SEAL GASKET Figure 7-9. Front Enclosure Gaskets Technical Service Manual 7 - 17 Plum 360 Infuser...
8. Join the front enclosure, rear enclosure, and main chassis in the exact reverse order of separation. 9. Reinstall the battery and connect the infuser to AC power. 10. Perform the PVT in Section 5.5 to verify successful pole clamp replacement. Technical Service Manual 7 - 19 Plum 360 Infuser...
2. Remove the battery as described in 3. Separate the front enclosure, rear enclosure, and main chassis as described Section 7.2.8. 4. Remove the rear enclosure gasket from the rear enclosure. Remove the power supply gasket. Plum 360 Infuser 7 - 20 Technical Service Manual...
Replacement of the minipole assembly is routine maintenance and no verification procedure is normally required. However, if there is a possibility the infuser may have Section 5.5. been damaged during the procedure perform the PVT as described in Technical Service Manual 7 - 21 Plum 360 Infuser...
Page 199
SECTION 7 REPLACEABLE PARTS AND REPAIRS CLUTCH HOUSING CLUTCH SPRING BAG HANGER POLE CLAMP EXTRUSION HAIRPIN CLIP Figure 7-11. Minipole Assembly Plum 360 Infuser 7 - 22 Technical Service Manual...
SCREW, 4-40 x 1/4 HEX HEAD SLOTTED, WITH WASHER MECHANISM CHASSIS BUMPER CPU/DRIVER CABLE MOTOR POWER 6-32 x 3/8 CABLE HEX HEAD SCREW WITH WASHER MECHANISM ASSEMBLY Figure 7-12. Main Chassis Assembly Components Technical Service Manual 7 - 23 Plum 360 Infuser...
8. Join the front enclosure, rear enclosure, and main chassis in the exact reverse order of separation. 9. Reinstall the battery and connect the infuser to AC power. Section 5.5 10. Perform the PVT in to verify successful power supply PWA replacement. Plum 360 Infuser 7 - 24 Technical Service Manual...
10. Join the front enclosure, rear enclosure, and main chassis in the exact reverse order of separation. 11. Reinstall the battery and connect the infuser to AC power. 12. Perform the PVT in Section 5.5 to verify successful keypad replacement. Technical Service Manual 7 - 25 Plum 360 Infuser...
7. Join the front enclosure, rear enclosure, and main chassis in the exact reverse order of separation. 8. Reinstall the battery and connect the infuser to AC power. 9. Perform the PVT in Section 5.5 to verify successful display assembly replacement. Plum 360 Infuser 7 - 26 Technical Service Manual...
Page 204
.20 X .18 X 7.15 LG DISPLAY CONDUCTIVE GASKET CONDUCTIVE 0.20 X 0.18 X 1.65 LG (3) GASKET .20 X .28 X 3.60 LG KEYPAD Figure 7-13. Keypad, Display, CPU PWA, and Piezo Alarm Technical Service Manual 7 - 27 Plum 360 Infuser...
Note: The cable’s black or red edge (pin 1) should be pointed up. Figure 7-14. Front and Rear View of CPU/Driver Cable Routed Through Main Chassis Plum 360 Infuser 7 - 28 Technical Service Manual...
Page 206
Receptacle pins Figure 7-17. Male Receptacle Header Socket on the CPU PWA Note: Verify that the pins inside the male receptacle socket housing on the CPU PWA are not damaged or bent. Technical Service Manual 7 - 29 Plum 360 Infuser...
Page 207
Figure 7-19. Side View of the CPU/Driver Cable Female Connector Firmly Seated and Locked to the CPU PWA Male Connector 13. Install the replacement CPU PWA. 14. Reinstall the display assembly. Plum 360 Infuser 7 - 30 Technical Service Manual...
Page 208
Figure 7-21. J11 Connector on the Infusion Mechanism Driver PWA Connector is seated Locking pin is and straight firmly locked in place Infusion mechanism Figure 7-22. CPU/Driver Cable Connected to the Infusion Mechanism Driver PWA Technical Service Manual 7 - 31 Plum 360 Infuser...
Page 209
Figure 7-24. Power Cable and Battery Cable Tie-wrapped 22. Connect the Piezo alarm cable. 23. Reinstall the power supply and display in the exact reverse order. 24. Install the front and rear enclosures. Plum 360 Infuser 7 - 32 Technical Service Manual...
Page 210
27. Perform the PVT in to verify successful CPU PWA replacement. MAIN CHASIS 4-40 x 3/16 SLOTTED HEX HEAD SCREW WITH WASHER (2) CPU INSULATOR CPU PWA Figure 7-25. CPU PWA Replacement Assembly Technical Service Manual 7 - 33 Plum 360 Infuser...
9. Join the front enclosure, rear enclosure, and main chassis in the exact reverse order of separation. 10. Reinstall the battery and connect the infuser to AC power. 11. Perform the PVT in Section 5.5 to verify successful piezo alarm assembly replacement. Plum 360 Infuser 7 - 34 Technical Service Manual...
10. Join the front enclosure, rear enclosure, and main chassis in the exact reverse order of separation. 11. Reinstall the battery and connect the infuser to AC power. Section 5.5 12. Perform the PVT in to verify successful infusion mechanism assembly replacement. Technical Service Manual 7 - 35 Plum 360 Infuser...
Page 213
SCREW, 4-40 x 1/4 HEX HEAD MAIN CHASSIS SLOTTED, WITH WASHER MECHANISM CHASSIS BUMPER CPU/DRIVER CABLE MOTOR POWER 6-32 x 3/8 CABLE HEX HEAD SCREW WITH WASHER MECHANISM ASSEMBLY Figure 7-26. Infusion Mechanism Assembly Replacement Plum 360 Infuser 7 - 36 Technical Service Manual...
Note: Installation of gasket tape to cover the opening on the infusion mechanism chassis helps protect the Switch PWA from being exposed to fluid ingress. The required tool for this procedure is the Hospira alignment fixture (P/N 519-97225-001). The replacement part for this procedure is:...
If the fluid shield tab is bent, contact Hospira for a fluid shield assembly replacement. Recommended tools for this procedure are a medium size flat-blade screwdriver and long needle nose pliers.
Page 216
CAUTION: Use extreme caution when installing or replacing the fluid shield. Ensure that the fluid shield is properly aligned with the infusion mechanism assembly pins. Technical Service Manual 7 - 39 Plum 360 Infuser...
Page 217
If the regulator closer is not seated properly, discontinue use and for technical assistance. Normal Tab Bent Tab Figure 7-30. Close-up Views of Normal (Left) and Bent (Right) Fluid Shield Tab Plum 360 Infuser 7 - 40 Technical Service Manual...
Page 218
A pump with an unseated regulator closer may pass the Unrestricted Flow Test. MECHANISM ASSEMBLY GASKET, .72 IN. (2) FLUID SHIELD GASKET, 1.09 IN. CASSETTE DOOR Figure 7-31. Fluid Shield Replacement Technical Service Manual 7 - 41 Plum 360 Infuser...
Page 219
LEVER COVER MECHANISM ASSEMBLY CASSETTE DOOR DOOR HANDLE 3/32 PUSH-ON RETAINING RING 4-40 x 3/8 HEX HEAD SCREW DOOR PIVOT CAP WITH WASHER Figure 7-32. Cassette Door and Door Handle Replacement Plum 360 Infuser 7 - 42 Technical Service Manual...
14. Join the front enclosure, rear enclosure, and main chassis in the exact reverse order of separation. 15. Reinstall the battery and connect the infuser to AC power. 16. Perform the PVT in Section 5.5 to verify successful door handle assembly replacement. Technical Service Manual 7 - 43 Plum 360 Infuser...
Page 221
SECTION 7 REPLACEABLE PARTS AND REPAIRS This page intentionally left blank. Plum 360 Infuser 7 - 44 Technical Service Manual...
Section 8 SPECIFICATIONS The following specifications apply to the Plum 360 Infuser. PHYSICAL: Dimensions: Approximately 8 H x 8 W x 6 D inches (20 cm H x 20 cm W x 15 cm D) (excluding pole clamp extrusion and power cord storage) Mass: Approximately 10 lbs (4.5 kilograms) with battery...
Page 223
Note 4: The optimal relative humidity for storage or operation is 25% to 85%. For short durations (up to 2 weeks), operation or storage at a relative humidity in the range of 10% to 90% is permissible. Plum 360 Infuser 8 - 2 Technical Service Manual...
Page 224
Ethernet LAN: DHCP assigned IP address, Subnet Mask, Gateway, DNS1, and DNS2 RoHS: The infuser meets the requirements of Directive 2011/65/EU on the restriction of the use of certain hazardous substances in electrical and electronic equipment. Technical Service Manual 8 - 3 Plum 360 Infuser...
Page 225
SECTION 8 SPECIFICATIONS This page intentionally left blank. Plum 360 Infuser 8 - 4 Technical Service Manual...
Section 9 Drawings in are provided as information only, and may not exactly reflect current product configuration. Note: PWA schematic drawings are available from Hospira upon request. Table 9-1. Drawings Figure Number Title Illustrated Parts Breakdown (3 Sheets)
Page 227
Tip, Shaft, Pole Clamp Section 7.2.11.2 Extrusion, with Rubber Pad, Pole Clamp Section 7.2.11.1 Retainer, AC Power Cord Section 7.2.5 Cordset, AC Power, Hospital Grade Section 7.2.5 Strap, Velcro Section 7.2.5 Plum 360 Infuser 9 - 2 Technical Service Manual...
Section 7.2.13.6 Tape, 3M 4955, .08 in. Thk., .25 in. W, White As applicable Tape, 3M 850, 1.9 MIL, .25 in. W, Clear As applicable Gasket, Fluid Shield, .72 in. Section 7.2.13.8 Technical Service Manual 9 - 3 Plum 360 Infuser...
Page 229
Rivet, .125, Christmas Tree, Black As applicable Screw, 4-40 x 3/8, Button Head, SS As applicable Screw, 4-40 x 5/8, Pan Head, Phillips Section 7.2.6 Screw, Jack, 4-40 x 7/16 Section 7.2.5 Plum 360 Infuser 9 - 4 Technical Service Manual...
Page 230
As applicable Insulator, Secondary As applicable Gasket, Peripheral Housing As applicable Lens, LED Section 7.2.7.2 Cover, USB Section 7.2.7 Seal, Round, Purple Section 7.2.7.3 Bolt, 4-40 x 1/4, Hex Head As applicable Technical Service Manual 9 - 5 Plum 360 Infuser...
Page 231
As applicable Screw, 6-32 x 5/8, Pan Head, Phillips, Square Cone As applicable Screw, 6-32 x 1/4, Pan Head, Phillips, SS As applicable Washer, Split Lock, #4, Regular, SS As applicable Plum 360 Infuser 9 - 6 Technical Service Manual...
Page 232
HOSPIRA, INC. Figure 9-1. Illustrated Parts Breakdown Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 3 Technical Service Manual 9 - 7 Plum 360 Infuser...
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 8 Technical Service Manual...
Page 234
HOSPIRA, INC. Figure 9-1. Illustrated Parts Breakdown Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 2 of 3 Technical Service Manual 9 - 9 Plum 360 Infuser...
Page 235
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 10 Technical Service Manual...
Page 236
112 113 114 HOSPIRA, INC. Figure 9-1. Illustrated Parts Breakdown Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 3 of 3 Technical Service Manual 9 - 11 Plum 360 Infuser...
Page 237
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 12 Technical Service Manual...
Page 238
HOSPIRA, INC. Figure 9-2. Front Enclosure, Rear Enclosure, Main Chassis, and CE Module Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 1 Technical Service Manual 9 - 13 Plum 360 Infuser...
Page 239
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 14 Technical Service Manual...
Page 240
HOSPIRA, INC. Figure 9-3. Front Enclosure Assembly Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 1 Technical Service Manual 9 - 15 Plum 360 Infuser...
Page 241
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 16 Technical Service Manual...
Page 242
HOSPIRA, INC. Figure 9-4. Rear Enclosure Assembly Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 1 Technical Service Manual 9 - 17 Plum 360 Infuser...
Page 243
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 18 Technical Service Manual...
Page 244
HOSPIRA, INC. Figure 9-5. CE Module Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 1 Technical Service Manual 9 - 19 Plum 360 Infuser...
Page 245
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 20 Technical Service Manual...
Page 246
HOSPIRA, INC. Figure 9-6. Main Chassis Assembly Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 2 Technical Service Manual 9 - 21 Plum 360 Infuser...
Page 247
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 22 Technical Service Manual...
Page 248
HOSPIRA, INC. Figure 9-6. Main Chassis Assembly Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 2 of 2 Technical Service Manual 9 - 23 Plum 360 Infuser...
Page 249
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 24 Technical Service Manual...
Page 250
HOSPIRA, INC. Figure 9-7. AC Power Cord Assembly and Battery Assembly Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 1 Technical Service Manual 9 - 25 Plum 360 Infuser...
Page 251
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 26 Technical Service Manual...
Page 252
HOSPIRA, INC. Figure 9-8. Antenna PWA Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 1 Technical Service Manual 9 - 27 Plum 360 Infuser...
Page 253
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 28 Technical Service Manual...
Page 254
HOSPIRA, INC. Figure 9-9. CPU PWA and Main Chassis Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 1 Technical Service Manual 9 - 29 Plum 360 Infuser...
Page 255
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 30 Technical Service Manual...
Page 256
HOSPIRA, INC. Figure 9-10. Infusion Mechanism Assembly Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 1 Technical Service Manual 9 - 31 Plum 360 Infuser...
Page 257
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 32 Technical Service Manual...
Page 258
HOSPIRA, INC. Figure 9-11. Minipole Assembly Rev. N/A DRAWING NO. NOT APPLICABLE Sheet 1 of 1 Technical Service Manual 9 - 33 Plum 360 Infuser...
Page 259
SECTION 9 DRAWINGS This page intentionally left blank. Plum 360 Infuser 9 - 34 Technical Service Manual...
Page 260
Table A-1. Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The Plum 360 (List numbers 30010, 30011, 30012) is intended for use in the electromagnetic environment specified below. The customer or the user of the Plum 360 should assure that it is used in such an environment.
Page 261
70% U (30% dip in U 70% U (30% dip in U IEC 61000-4-11 for 25 cycles for 25 cycles The Plum 360 includes an internal <5% U (>95% dip in U <5% U (>95% dip in U battery backup that...
Page 262
Level Immunity Guidance Portable and mobile communications equipment should be used no closer to any part of the Plum 360, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Conducted RF...
Page 263
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Plum 360 is used exceeds the applicable RF compliance level above, the Plum 360 should be observed to verify normal operation.
Page 264
RECOMMENDED SEPARATION DISTANCES FOR COMMUNICATIONS EQUIPMENT The Plum 360 (List numbers 30010, 30011, and 30012) is intended for use in an electromagnetic environment in which radiated disturbances are controlled. The customer or user of the Plum 360 can help prevent electromagnetic interference by maintaining...
Page 265
APPENDIX This page intentionally left blank. Plum 360 Infuser A - 6 Technical Service Manual...
Page 266
1-800-241-4002 For additional services and technical training courses, visit the website: www.hospira.com For technical assistance and services outside the United States, contact the local Hospira sales office WARNING: EXPLOSION HAZARD EXISTS IF THE INFUSER IS USED IN THE PRESENCE OF FLAMMABLE SUBSTANCES.
Page 267
Caution Warning Mains supply equipment Protection against vertically IPX2 using protective earth falling water drops Fragile, Handle with Care Temperature limitation This Way Up Keep dry Device includes an RF transmitter Dangerous Voltage that complies with IEEE 802.11 a/b/g/n Mandatory Action Follow Instructions For Use Alarm Volume Control Nurse Call Interface Port...
Need help?
Do you have a question about the Plum 360 and is the answer not in the manual?
Questions and answers