Troubleshooting guide
The following troubleshooting guide describes a number of faults and events which may occur when you use your
wheelchair, together with suggested remedies. Note that this guide cannot describe all the problems and events which
may occur and you should always contact your service contact or Permobil in case of doubt.
EVENT
Wheelchair cannot be started.
Wheelchair cannot be driven.
Battery voltage indicator on the control panel
flashes at 2.5 second intervals and the wheel-
chair cannot be driven.
Battery voltage indicator on the control panel
flashes quickly and the wheelchair cannot be
driven.
Wheelchair stops while being driven.
Wheelchair can only be driven at reduced
speed. Applies to wheelchairs with an electric
seat lift.
118
POSSIBLE CAUSE
Batteries discharged.
Main fuse switched to OFF position after, for
example, battery replacement.
Main fuse triggered.
Battery charger connected.
Brake release activated.
Wheelchair locked with the security key.
Electronics' power saving position is
activated.
Electronics fault.
Main fuse triggered.
Seat lift or seat angle raised too high.
REMEDY
Charge the batteries.
Reset the main fuse. See page 107.
See page 107.
Stop charging and disconnect the charging
cable from the wheelchair's charging socket.
Reset the brake release.
Unlock the wheelchair.
Switch off and switch on the power using the
start key on the control panel.
Contact service.
See page 107.
Lower the seat lift.